Fault
Cause
Solution
Instruments
do not
hold
•
Maximum permitted speed
for instrument in use is
exceeded
•
Instrument shaft worn
•
Instrument shaft bent
•
Instrument not correctly
inserted in clamping
mechanism
•
Cream, ointment, or similar
on instrument shaft
•
To much work pressure on
instrument
•
Pulling movement while
working
•
Only work with the
maximum permissible
speed for the instruments
used
•
Check the instrument shaft
and use a new instrument
if necessary
•
Replace instrument
without fail
•
Insert instrument until stop
in clamping mechanism
•
Clean instrument. Work
with small instruments
until larger instruments
start to clamp securely
again. During operation,
always make sure that the
handpiece tip and the
instruments are clean.
•
Reduce pressure
•
Avoid pulling movements
Speed only
possible up to
10,000 rpm.
Device
automatically
switches back
to 10,000 rpm.
Insufficient spray solution in tank.
The device switches back to
10,000 rpm to prevent the
handpiece from getting warm.
Refill spray liquid
Handpiece
vibrates;
Handpiece is
loud with used
instruments
•
Maximum permitted speed
for instrument in use is
exceeded,
•
Instrument defective (bent,
shaft worn, etc.)
•
Only work with the
maximum permissible
speed for the instruments
used
•
Use new instruments.
Push button on
handpiece
blocked or
is difficult to
activate
Handpiece or handpiece button
soiled
Clean button
If one of these faults still persists despite having followed this information, please send the
unit to the customer service team.
In addition, please get in touch with the manufacturer for each fault / each malfunction in
cases of doubt.