Operating Manual BaehrTec S2000
English– V101 – Status 07/2015
Page 31/52
6.
Take the handpiece in one hand and insert an instrument with a shank diameter 2.35
mm as deep as possible into the opening DIN instruments
(55)
. The head of the
instrument must not be larger than that shown on the front panel (max. 12 mm).
Never use instruments with an oily, worn or bent shaft. Otherwise, it cannot
be guaranteed that your instrument can be held firmly in the handpiece!
Caution – risk of injury! Never
attempt to insert or remove instruments
into/ out of the opening
(55)
when the handpiece motor is running.
Instruments can only be changed when the handpiece motor is switched off.
The instrument heads indicated on the front foil are intended to help you
identify the maximum speed for your instruments. However, before you work
with an instrument, be sure to follow the manufacturer’s instructions on the
permitted maximum speed. The maximum speed specified there must not
be exceeded under any circumstances. This poses a danger of injury and
the danger that your device will be damaged.
7.
Now select the permitted maximum speed for the instrument you currently want to
work with. The instrument heads depicted (max. diameter) and the associated speed
should help you to quickly and safely find the permitted maximum speed for your
instruments.
The permitted maximum speed must not be exceeded under any
circumstances
; otherwise the instrument or the handpiece may be damaged. It can
also cause injury due to broken instrument heads. You can find out the correct
permissible maximum speed for the instrument that you are using by comparing the
instrument head diameter with the instrument heads shown on the front foil. Once you
have found a match, press the illustrated instrument head that you have found to be
correct. The permitted
maximum
speed is now set. The speed is shown in the display.
Lower speeds are generally permitted for all of the instruments.
WARNING: This is a speed recommendation. Please refer to the data sheet of the
instrument manufacturer to find out the permitted maximum speed of the instrument.
The permitted maximum speed must not be exceeded under any circumstances;
otherwise the instruments or the handpiece may be damaged. It can also cause injury
due to broken instruments.
The speed ranges 6,000 and 10,000 rpm for the instruments with 12 mm
diameter (abrasive caps/DiaTWISTER) are not designed to be used to
remove calluses, but rather
to smoothen
horny skin (with little pressure).
To remove calluses
select instruments (abrasive caps/DiaTWISTER) with 10
mm diameter for the speed ranges 15,000 or 20,000 rpm.
8.
You can now turn on the handpiece motor using the push button
(13)
on the controller
or the push button
(52b)
on the handpiece. The arrow in the display will now be filled
blue. The instrument will now turn clockwise at the speed you have selected. You can
now change the motor speed of the handpiece by pressing the buttons
(5)
– (12)
,
however, you must not exceed the highest permissible speed for the instrument being
used. The selected speed is indicated on the display. You can change the instrument
speed in steps of 1,000 with the buttons
(14)
and
(15)
.
Please note that by changing the rotation speed in steps of 1,000 (button
(14)
or button
(15)
), also the maximal permissible instrument size when you have
reached to defined maximum rotation speed using the buttons
(5)
-
(12)
. This
is indicated by the blue LED in the instrument area
(5a)
-
(12a)
. The actual
speed is shown on the display
(20)
.
E.g.:
The speed is set to 6,000 rpm. Now press the button
(14)
until you reach the
next defined rotation speed (10,000 rpm = button
(6)
) by pressing the buttons
(5)
-
(12)
. Now the LED for 10,000 rpm lights up automatically
(6a)
). This