Calibration
CALIBRATION
The Hydraulic System Analyzer is calibrated at the factory, but you may need to reset zero for the pressure and flow rate
sensors occasionally Calibration parameters are stored within the electronic interface and not the PC software When the
software application is launched, it automatically reads the calibration parameters from the device Therefore, owners of
multiple units can use the same PC without worrying about calibration files
Resetting Zero for the Pressure Transducer
1 Select
Tools > Calibration
from the main menu
Figure 6: Calibration screen
2 Select the
Pressure
or
Flow
radio button at the top of the calibration screen
3 Make sure the pressure/flow of the Hydraulic System Analyzer is at zero and click
Read
4 If the
Calibrated Value
reading under the
Units
column is not around zero, click
Set Zero
to recalibrate the sensor's zero
output and write the value to the Hydraulic System Analyzer
Resetting Factory Calibration Defaults
Click
Restore Factory Settings