Badger Meter ER-10 Installation & Operation Manual Download Page 7

7

OP

INDUSTRIAL

TURBO

DISC

If replacement of your sensor pickup on  your meter is

required, please follow the following steps:

INDUSTRIAL TURBO METER

1. Remove the front cover on the ER-10 unit and

disconnect all wiring.

2. Remove meter head bolts and lift meter head

assembly from housing.

3. Remove retaining ring which retains the accessory

unit to the head.

4. Loosen the side seal screw on the accessory

adapter, twist 90 degrees

 

and pull entire assembly

unit from the meter head.

5. Twist drop pipe in counterclockwise direction to

remove it and the ER-10 unit from the adapter
assembly.

6. Obtain new adapter assembly and reassemble to

drop pipe and ER-10 unit.

7. Reverse the balance of the above steps.

DISC METER

1. Remove the front cover on the ER-10 unit and

disconnect sensor wiring.

2. Loosen the side seal screw on the assembly

adapter, twist 90 degrees

 and 

lift entire accessory

unit off bare meter.

3. Pull the reed switch pickup assembly and pad from

adapter.

4. Replace pickup and pad in adapter, feeding wires up

through the drop pipe.

5. Reposition entire assembly on meter.

6. Rewire sensor to ER-10 unit.

OP METER

1. Remove the front cover on the ER-10 unit and

disconnect sensor wiring.

2. Remove the back plate on the meter to expose the

pickup assembly.

3. Remove the reed switch pickup assembly and

replace.

4. Feed wires up through the drop pipe.

5. Reassembly the back plate in position.

6. Rewire sensor wires to ER-10 unit.

METER MOUNT SURFACING

(ER-10)

Summary of Contents for ER-10

Page 1: ...ins Lithium Batteries Read This Manual Before Attempting Any Installation Wiring Or Operation ER 10 ER 10H For Portable Large Meter Tester Only Models ER 10 ER 10H Digital Resettable Totalizers and Di...

Page 2: ...nical support group at Badger Meter Inc Technical Support 1 800 456 5023 Specifications The ER 10 and ER 10H are external or battery powered electronic registers that display rate of flow and total fl...

Page 3: ...coming from the transmitter to the corresponding terminal number for the ER Remote models To connect a generic reed switch to the ER 10 connect one of the wires to terminal 1 Connect the remaining wir...

Page 4: ...he rate scale factor The lower case d appears on the right of the display when it is time to enter the decimal point position for the rate scaler NOTE This decimal point is used for the rate scaler on...

Page 5: ...te seconds would equal 60 If you wish to read flow in units per hour seconds would equal 3600 Transmitter Pulses per Unit is the number from the chart to the right or the Industrial Tech Brief ITB for...

Page 6: ...per unit of measure for your particular meter transmitter combination and then using this value as the transmitter pulse output value when calculating the counter and time base values on page 4 The pr...

Page 7: ...emble to drop pipe and ER 10 unit 7 Reverse the balance of the above steps DISC METER 1 Remove the front cover on the ER 10 unit and disconnect sensor wiring 2 Loosen the side seal screw on the assemb...

Page 8: ...the extent an outstanding contractual obligation exists BadgerMeter Inc P O Box 245036 Milwaukee WI 53224 9536 Telephone 414 355 0400 877 243 1010 Fax 414 355 7499 866 613 9305 www badgermeter com Cop...

Reviews: