Badger Meter ER-10 Installation & Operation Manual Download Page 2

General Information

POWER

Internal Battery: 3V, Lithium
Battery Life Expectancy: 5 years
Replacement Part #: 65765-001
External DC Power Source(10-30VDC)
Max Current Draw = 30mA

BACKLIGHT

Requires 10-30VDC power input (max. current
draw= 30mA)
(Derate operating temperature 1°C/Volt
above 17VDC)
Reverse polarity protected

PHYSICAL

Operating Temperatures:

32°F (0°C) to 130°F (55°C)

Storage Temperature:

-4°F (-20°C) to 160°F (70°C)

Operating Humidity: 60% Non-condensing
Weight: 2.2 oz. net
Display Size: .43" high
Front Panel Rating: NEMA 4X when mounted
with gasket provided
Case Material: Cycolac X-17

TOTALIZER

Digits: 8
Scaler: 0.0001 - 99.999
Decimal Point: 5 positions, programmable

2

Scope of this manual

This manual contains information concerning the installation,
operation and maintenance of the Badger

®

 ER-10 and ER-

10H electronic registers. To ensure proper performance, the
instructions given in this manual should be thoroughly
understood. Retain the manual in a readily accessible location
for future reference.

Installation, wiring and programming of either unit is fairly
simple and straight forward. This manual is designed so as to
provide you with a step by step guide for this purpose.

Examples are provided only to facilitate programming.
Your specific application will most likely require a different
set of numbers for proper programming.

The troubleshooting section attempts to illustrate the
most common problems that can be encountered, their
most likely cause and the recommended solution. If a
problem persists, please contact our technical support
group at:

Badger Meter, Inc.
Technical Support

1-800-456-5023

Specifications

The ER-10 and ER-10H are external or battery powered
electronic registers that display rate of flow and total flow.
They have independent programmable scale factors for
rate and totalization, allowing you to program these dis-
played values in different but meaningful engineering
units, such as gallons per minute and total gallons.

The LCD display with 8 digits for total, and 4 digits plus
legend for rate, provides easy viewing at a glance.  For
conditions where ambient light is poor, the display can be

backlit by connecting an external DC (10-30 VDC) power
supply. A single unit can accept NPN or dry contact inputs
for low or high speed applications.

The 

ER-10 and ER-10H

 models are powered by one 3V

Lithium battery.

Setup on either model is quick and easy as the two front
panel keys are used to scroll through and preset values
in all program mode choices.

RATE INDICATOR

Digits: 4/5, (4 calculated, 5 displayed with
fixed 0 in LSD)
Scaler Range: .001to 9999
Decimal Point: 5 positions, programmable
Accuracy: +/-0.2%
Update Time: .7 seconds
Zero Time: 10 seconds

DC COMMON (Terminal 1)
COUNT INPUT (Terminal 2)

Speed: 0 to 100 hz
Min. Low Time: 3.3 mS
Min. High Time: 6.7 mS
Impedance: 101K ohm
Voltage  Thresholds:

Low: 0 to 0.4 VDC
High: 2.0 to 28 VDC
Max: High 28 VDC

RESET INPUT (Terminal 4)

Resets totalizer to zero when connected to DC
common.
Min. Low Time: 0.25 to 1.0 sec. (maintained)
The required pulse width varies with count
speed, scale factor and number of digits
displayed.
Voltage  Thresholds:

Low: 0 to 0.4 VDC
High: 2.0 to 28 VDC

PROGRAM ENABLE INPUT (Terminal 5)

Operation: Level sensitive (maintained)

COUNT ACCURACY

100% when operated within specifications

Summary of Contents for ER-10

Page 1: ...ins Lithium Batteries Read This Manual Before Attempting Any Installation Wiring Or Operation ER 10 ER 10H For Portable Large Meter Tester Only Models ER 10 ER 10H Digital Resettable Totalizers and Di...

Page 2: ...nical support group at Badger Meter Inc Technical Support 1 800 456 5023 Specifications The ER 10 and ER 10H are external or battery powered electronic registers that display rate of flow and total fl...

Page 3: ...coming from the transmitter to the corresponding terminal number for the ER Remote models To connect a generic reed switch to the ER 10 connect one of the wires to terminal 1 Connect the remaining wir...

Page 4: ...he rate scale factor The lower case d appears on the right of the display when it is time to enter the decimal point position for the rate scaler NOTE This decimal point is used for the rate scaler on...

Page 5: ...te seconds would equal 60 If you wish to read flow in units per hour seconds would equal 3600 Transmitter Pulses per Unit is the number from the chart to the right or the Industrial Tech Brief ITB for...

Page 6: ...per unit of measure for your particular meter transmitter combination and then using this value as the transmitter pulse output value when calculating the counter and time base values on page 4 The pr...

Page 7: ...emble to drop pipe and ER 10 unit 7 Reverse the balance of the above steps DISC METER 1 Remove the front cover on the ER 10 unit and disconnect sensor wiring 2 Loosen the side seal screw on the assemb...

Page 8: ...the extent an outstanding contractual obligation exists BadgerMeter Inc P O Box 245036 Milwaukee WI 53224 9536 Telephone 414 355 0400 877 243 1010 Fax 414 355 7499 866 613 9305 www badgermeter com Cop...

Reviews: