ID
Parameter
Value
Notes
3 10
Reporting day of the month
-- -- 31
INF
3 11
Tariff 1 configuration
Example of Tariff 1, when
T1-T2 is < 10 0 °C:
Possible setting:
One of measured parameters, 1 or 2
pulse input (if it is configured as an
input), one of the temperatures, or
temperature differential
1-2
L1 10 0 °C
INF
MAX
when >10 0 °C:
1-2
L1 10 0 °C
INF
MIN
when within the range from 10 0 to
40 0°C (changes every 1 second):
1-2
L1 10 0 °C
INF
MIN
1-2
L1 40 0 °C
INF
MAX
when the time interval is set in hours
(07-23 h):
1-2
L1 07-32
h
INF
MAX
3 12
Tariff 2 configuration
Similarly to Tariff 1, only “L1” changes to
“L2”
Similarly to tariff 1
3 14
2
nd
pulse input/output
configuration
Similarly to the 1
st
pulse input/output,
only “1” changes to “2”
Similarly to the 1
st
pulse input/output
3 15
Wire interface M-Bus 1 address
1
buSA 1
INF
3 16
Wire interface M-Bus 1 speed
1
2400E bPS
INF
Bits per second
“E” – parity Even
3 17
Wire interface M-Bus 2 address
2
buSA 1
INF
When a second wire interface is also
included
Operating Procedure
Page 23
November 2020
HYB-UM-03193-EN-02