Electrical Wiring
If the meter is to be powered from an external 230V AC or 24V AC/DC source, the cable of the meter intended for the purpose
and respectively marked is connected to the respective source (see
"Transportation and Storage" on page 30
If the meter is completed with wire interfaces or the
Pulse Input/Output
function, the cables intended for the purpose and
respectively marked are connected to the respective external appliance (see
"Transportation and Storage" on page 30
Mounting the Calculator
Mount the electronic unit (calculator) of the meter in a heated room The temperature of the working environment should not
be higher than 131° F (55° C) It may not be exposed to direct sunlight
No special requirements are established for the free space around the meter It is important that nearby installations or
structures do not rest against the housing of the meter, do not hinder the laying of cables and reading of data on the display
Install the meter at a safe distance from other devices emitting heat or strong electromagnetic field (in order to prevent the
disturbance of its working environment conditions)
The electronic unit is mounted on an auxiliary holder It can be oriented in the required direction at an angle of each 90:
The possible ways of the mounting of the electronic unit (auxiliary holder):
Direct mounting on the housing of the flow sensor, by turning each 90° only when the flow temperature does not exceed
194° F (90 °C):
On a wall:
In the electrical equipment cabinet, on a standard DIN rail:
Installation Procedure