Badger Meter Blancett B2800XP User Manual Download Page 12

ADDITONAL SCALING PAOAMETEOS

x

x

Figure 5:  4…20 mA programming menu

OTEE:

N

The programming instructions in this section are only available for loop powered units  Battery powered units do not 

include these programming parameters 

Flow 4 mA Setting

Zero is the default flow rate at the 4 mA setting  If the current selection is correct, press 

ENTEO

 to advance to the 

Flow 20 mA

 

parameter  To change the setting, press 

UP

 to increment the flashing digit and press 

OIGHT

 to move to the next digit  Press 

ENTEO

 to save and advance to the 

Flow 20 mA

 parameter 

Flow 20 mA Setting

The turbine meter’s maximum flow rate is the default value for the 20 mA setting  If the current selection is correct, press 

ENTEO

 to advance to the 

4…20 mA Calibration

 parameter  To change the setting, press 

UP

 to increment the flashing digit and 

press 

OIGHT

 to move to the next digit  Press 

ENTEO

 to save and advance to the 

4…20 mA Calibration

 parameter  

4…20 mA Calibration

The 

4…20 mA Calibration

 parameter is the fine adjustment of the 4…20 mA output  The 4 mA setting is typically between 

35…50  To set the 4 mA value, connect an ammeter in series with the loop power supply  At the 

4MA OUT

 prompt, press 

UP

 to 

increment and 

OIGHT

 to decrement the 4 mA value to until a steady 4 mA reading is reached on the ammeter  Press 

ENTEO

 to 

lock in this value and advance to 

20MA OUT

  The 20 mA adjustment is performed using the same procedure as the  

4 mA adjustment 

4…20 mA Test

The monitor contains a diagnostic routine that allows the simulation of mA values between 4…20 to check output tracking  

At the 

4-20TEST

 prompt press 

UP

 to increment the simulated mA output or 

OIGHT

 to decrement the value in increments of  

1 mA  The ammeter should track the simulated mA output  If a 4…20 mA test in not necessary, press 

ENTEO

 once to escape 

the testing at any time 

Pulse Out

Freq. In

4-20 mA

10…30V DC

 

1…5V DC 

1

2

3

4

5

6

+          -

+         -

+          -

JP2

Mag

Input

Pulse

Inuput

TB1

1 2 3 4 5

J2

Keypad

Connector

DC

Figure 6:  Typical ammeter connection

Additonal Scaling Parameters

Page 12 

March 2017 

DSY-PM-00279-EN-03

Summary of Contents for Blancett B2800XP

Page 1: ...Flow Monitor B2800XP Explosion Proof Flow Monitor Advanced DSY PM 00279 EN 03 March 2017 User Manual...

Page 2: ...Mode 11 Additonal Scaling Parameters 12 Flow 4 mA Setting 12 Flow 20 mA Setting 12 4 20 mA Calibration 12 4 20 mA Test 12 Additional Input Options 13 Battery Replacement 13 Maintenance 14 Additional...

Page 3: ...serious personal injury Indicates a hazardous situation which if not avoided could result in severe personal injury or death Indicates a hazardous situation which if not avoided is estimated to be ca...

Page 4: ...lectable units viewing in run mode and prompts for variables in program mode The display can show rate total or alternate between both rate and total MENU ENTER Programming Buttons Figure 1 B2800XP fl...

Page 5: ...amage to the cover threads and glass lens when glass lens cover is used 5 Pull wires into enclosure making certain they are long enough to make the required connections and to remove the instrument or...

Page 6: ...b Size Kit Number Meter Sizes 1 2 in NPT Hub B280 742 3 8 in 1 2 in 3 4 in with 1 2 in NPT End Fittings 1 in NPT Hub B280 737 All Sizes with 1 in NPT End Fittings and Larger Installation Kit P N B280...

Page 7: ...ion turn on the lock out feature to prevent unauthorized access or changing the meter s setup parameters MENU ENTER Figure 3 Front panel Buttons MENU Enters and exits PROGRAM mode UP Scrolls forward t...

Page 8: ...If the current setting is correct press ENTER to advance to the next parameter To change the current setting press UP or RIGHT to scroll to the bore size of your meter Press ENTER to save and advance...

Page 9: ...rams were chosen as the RATE UNT or TOTL UNT the monitor advances to the SP GRAV parameter Specific Gravity At the SP GRAV parameter press ENTER The first digit of the current specific gravity flashes...

Page 10: ...plitude of the pulse is dependent on the voltage level of the supply connected to the pulse output and is limited to a maximum 30V DC Linearization Linearization enhances the accuracy of the display T...

Page 11: ...password and return to the METER parameter The reset pasword is now required to reset the totals on the monitor NOTE N Entering a password in the PASSWORD screen and leaving the default password in t...

Page 12: ...4 20 mA Calibration The 4 20 mA Calibration parameter is the fine adjustment of the 4 20 mA output The 4 mA setting is typically between 35 50 To set the 4 mA value connect an ammeter in series with...

Page 13: ...Power Figure 7 Circuit board layouts BATTERY REPLACEMENT Battery powered monitors use a single 1 5V D size alkaline battery When replacement is necessary use a clean fresh battery for continued troub...

Page 14: ...ates some form of alternating signal that can be raw that is no format amplification or wave shaping can be done prior to transmission to the readout The output pulse rate is related proportionally to...

Page 15: ...t 2 25 Pulse Output Signals for Non linear Processes Few flow meters actually behave in a linear way There is always some uncertainty about the exact flow at a given reported input value For many comm...

Page 16: ...nitor pig tail This should stop the noise If not then try reorientating the meter to a new location Flow rate indicator bounces This usually indicates a weak signal Replace pickup and or check all con...

Page 17: ...ity 30 mV p p Over Voltage Protected 30V DC Frequency Measurement Accuracy 0 1 Temperature Drift 50 ppm C maximum Optional Analog Output Loop Powered Version Type 4 20 mA current loop Resolution 1 400...

Page 18: ...RT NUMBER INFORMATION NOTE Advanced B2800 monitors only The default is gallons per minute Program Level A Advanced B 28 A X X X X Power Option B Battery L Loop Mounting Style X Explosion proof Units o...

Page 19: ...INTENTIONAL BLANK PAGE Page 19 March 2017 DSY PM 00279 EN 03...

Page 20: ...Colonia Guadalupe Inn CP 01050 M xico DF M xico 52 55 5662 0882 Europe Eastern Europe Branch Office for Poland Latvia Lithuania Estonia Ukraine Belarus Badger Meter Europe ul Korfantego 6 44 193 Knur...

Reviews: