Password
The password setting restricts access to
PROGRAM
mode Initially, the password is set to all zeros and any user can modify
the paramter settings To change the password, press
ENTEO
at the
PASSWORD
prompt The first digit flashes Press
UP
to
increment the digit and
OIGHT
to advance to the next digit After entering all digits, press
ENTEO
to store the password and
advance to
RST PSWD
The new password is now required to enter
PROGRAM
mode With this password set, any user is able to
reset the stored totals on the monitor
OTEE:
N
This password will allow the operator to manually reset totals
Oeset Password
The reset password parameter restricts resetting the totals on the monitor The
PASSWORD
must also be set to restrict the total
reset Initially, the password is set to all zeros and any user can reset the stored totals on the monitor To change the password,
press
ENTEO
at the
RST PSWD
prompt The first digit flashes Press
UP
to increment the digit and
OIGHT
to advance to the next
digit After entering all digits, press
ENTEO
to store the password and return to the
METER
parameter The reset pasword is
now required to reset the totals on the monitor
OTEE:
N
Entering a password in the
PASSWORD
screen and leaving the default password in the
RST PSWD
screen allows for
total resets (not requiring a password), but restricts programming modification
Oeturn to Oun Mode
After entering all parameters, press
MENU
to return to
RUN
mode
Oun Mode
Oeset Total
To reset the monitor’s total display, in
RUN
mode, press
MENU
and
ENTEO
simultaneously until
TOTAL RST
starts to flash The
display returns to the
RUN
mode at the conclusion of the reset procedure
Store Total
The current total can be manually stored in the monitor’s flash memory This procedure may be needed prior to changing the
settings or replacing the battery Press and hold
ENTEO
for two seconds The display flashes
TOTALSVD
and then returns to
RUN
mode
Automatic Store Total
The monitor is equipped with a store total feature that works automatically, saving the current total to flash memory The
frequency of saves depends on the power supply option chosen
Battery Powered: Once per hour and just before a low battery condition turns the unit off
Loop Powered: Once every ten minutes
Programming
Page 11
March 2017
DSY-PM-00279-EN-03