Pur•Chek
™
Pro
Instruction 2100-9001
8
4 Operation
4.1 Battery Installation
The Pur•Chek™ Pro has, as an option, an internal rechargeable battery. If your Pur•Chek™ Pro is equipped
with the optional rechargeable battery, you must fi rst install and charge the battery prior to use. NOTE: The
Pur•Chek™ Pro can be operated with or without the battery using the supplied AC adapter.
To install the optional battery, remove the battery cover from the back of the unit by unscrewing the two Phillips
head screws as shown below.
OR OPENED CONTAINERS OF GASOLINE.
at least 18 inches above to floor.
Use this equipment in locations with mechanical ventilation that
provides at leat four air changes per hour or locate the equipment
D
A
TE
C
O
D
E
:
WARNING-TO REDUCE THE RISK OF FIRE:
DO NOT USE THIS EQUIPMENT IN THE VICINITY OF SPILLED
LISTED 6P03
Design Pressure: 300 psig
Refrigerant Type: R-12, R-134a
R
Miscellaneous Refrigerant Equipment
12 VDC, 1A; Design Pressure: 300 psig
CFC-12 (R-12) or HFC-134a (R-134a)
(R-134a).
For Accuracy, see Manual.
detect 2% or greater of air in HFC-R134a
to 98% purity. Also this equipment will
to Meet SAE J1771 to identify
Refrigerant Diagnostic Tool Design
Certified by Intertek Testing Services (ITS)
SAE J1771 CERTIFICATION
Battery Cover Screws
Inside the Pur•Chek™ Pro battery compartment, locate the male plug on the left side. Slide the nylon strap
around the battery. Insert the battery module into the compartment aligning the female connector of the battery
module with the male plug in the battery compartment. Replace the cover and snug the screws.
NOTE:
Charge the battery for a minimum of 4 hours with the supplied charger prior to fi rst use.
The battery cannot be charged with the standard power adapter.
To remove the battery, simply tug gently on the nylon strap, being sure to pull straight up, until the battery is
dislodged. The battery may be charged either inside of the unit or independent of the unit.