background image

Pur•Chek

 Pro

Instruction 2100-9001

15

 

  TO CHANGE
SETTINGS OR
REPLACE PAPER
ROLL PRESS SET
  FOR OTHER HELP
PRESS HELP

Figure 22.

Figure 23.

SET       HELP     DONE

SET       FEED     DONE

 

  TO FEED IN A 
NEW PAPER TOLL
PRESS FEED
  PRESS SET TO
ADJ LCD CONTRAST
OR ELEVATION

Figure 24.

 

     TO LOAD PAPER
1. PRESS FEED
2. INSERT PAPER
3. PRESS STOP
4. PRESS TEST TO
    TEST AND EXIT

FEED       STOP     TEST

Open the printer door and remove the old roll by tearing the paper as it enters the printer then pressing the 
“FEED” button shown in Figure 24 until the old roll exits the printer completely.  Insert the new paper roll from 
the underside as shown below:

CORRECT

INCORRECT

Press the “FEED” button shown in Figure 24 to automatically advance the paper through the printer.  When 
the paper appears on the top of the printer, press the “STOP button.  Press the “TEST” button to print a test 
message and verify that the paper is installed properly.  Slide the paper through the slot in the printer door and 
close the door.

5.6 Error Messages

In the unlikely event that an “Error” message is displayed on the screen, power off the unit, take it to a location 
outside of the shop environment where fresh air is available and turn the unit back on.  If the “Error” message 
reappears, refer to the help screens on the instrument or contact our service department for assistance.

5.7 Software Updates

From time to time, software updates may be made available to improve operating performance or add addi-
tional features.  Some updates will be provided at no charge to implement operating effi ciencies while others 
will be optional, paid upgrades, to add new refrigerants etc.  Many of the updates can be completed by the user 
however some will require the instrument to be returned to the factory for new gas calibrations.

The Pur•Chek™ Pro has a USB update port located within the battery compartment.  This port should not be 
used for any other purpose other than to install factory updates using the Bacharach factory USB drive.

R

Summary of Contents for PUR-CHECK PRO

Page 1: ...Product Leadership Training Service Reliability Instruction 2100 9001 Operation and Maintenance Rev 0 June 2010...

Page 2: ...LIEU OF I ANY AND ALL OTHER WARRANTIES EXPRESS OR IMPLIED INCLUDING WITHOUT LIMITATION THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE AND II ANY OBLI GATION LIABILITY...

Page 3: ...3 6 Back Panel Connections 7 3 7 Hard Shell Storage Carrying Case 7 4 Operation 8 4 1 Battery Installation 8 4 2 Turning On the Unit 9 4 3 Calibration 9 4 4 Vapor Sampling 10 4 5 Liquid Sampling 10 4...

Page 4: ...Pur Chek Pro Instruction 2100 9001 ii...

Page 5: ...rformed or adhered to could result in minor or moderate injury IMPORTANT The use of the word IMPORTANT in this manual calls attention to a procedure practice condition or the like which if not correct...

Page 6: ...tain or will not emit heavy loads of oil or liquid NEVER admit any sample into the instrument at pressures in excess of 500 psig NEVER obstruct the air intake sample exhaust or case vent ports of the...

Page 7: ...The instru ment utilizes non dispersive infrared NDIR technology to determine the weight concentrations of multiple refrigerant types Refrigerant purity is displayed on the LCD Screen The user must de...

Page 8: ...rared NDIR Refrigerant Sample Size 0 3 ounces 8 5 grams per sample Power External Power Supply 110 220V 5 60Hz 0 6A External Battery Charger 110 220V 50 60Hz 1 95A DC Supply Requirements 12 15 VDC 2 A...

Page 9: ...void the warranty Printer Door Air Intake Port Sample Filter Control Panel 3 2 Low Side Vapor Sample Hose The 6 foot 1 8 meter Low Side Vapor Sample Hose is constructed of a polyurethane tube A brass...

Page 10: ...AE Flare Nut 3 4 AC Power Adapter The Pur Chek Pro is powered via a 110 220 VAC 50 60 Hz power transformer This transformer is included with each unit and converts a standard 110 220 VAC 50 60 Hz wall...

Page 11: ...let port free and clear at all times Do not operate near open flame Sample Inlet Sample Outlet Battery Charge Port 12 VDC Power Input AC Adapter 3 7 Hard Shell Storage Carrying Case The hard shell sto...

Page 12: ...rant Type R 12 R 134a R Miscellaneous Refrigerant Equipment 12 VDC 1A Design Pressure 300 psig CFC 12 R 12 or HFC 134a R 134a R 134a For Accuracy see Manual detect 2 or greater of air in HFC R134a to...

Page 13: ...alibration takes 95 seconds Figure 4 and brings fresh air into the unit via an internal pump This fresh air purges any excess refrigerant from the unit and ensures accurate test results Calibration re...

Page 14: ...om pletely depressed 3 Connect the appropriate end of the hose to the instrument and the opposite end of the liquid sampling assembly to the High Side Liquid port of the system or cylinder Affix the o...

Page 15: ...lay the detail screen shown in Figure 11 If the blend or refrigerant mixture is not recognized the screen shown in Figure 12 will be displayed BACK MORE Figure 10 Figure 12 Figure 11 RESULTS R410A CON...

Page 16: ...procedure To exit the test press the EXIT button Figures 15 16 17 show sample printouts for various test results NOTE Care must be taken to leave a clean edge when tearing off the printed results Neu...

Page 17: ...he ELEV button will take you to the elevation screen shown in Figure 19 Use the UP DOWN buttons to adjust the unit to the elevation in your area Press the SAVE button when finished to return to the he...

Page 18: ...Oil contamination cannot be cleaned out of sample hoses due to the density of the oil restrictor If oil is visible in the Low Pressure Vapor hose replace the hose and restrictor with P N 2100 0010 Fo...

Page 19: ...lled properly Slide the paper through the slot in the printer door and close the door 5 6 Error Messages In the unlikely event that an Error message is displayed on the screen power off the unit take...

Page 20: ...2100 0010 Spare Low Pressure Sample Hose Filter 2100 0012 Spare Rechargeable Battery 2100 9001 Instruction Manual 7 Service Centers Service and replacement parts can be found by contacting any of the...

Page 21: ...Pur Chek Pro Instruction 2100 9001 17 NOTES...

Page 22: ...ers 621 Hunt Valley Circle New Kensington PA 15068 PH 724 334 5000 Toll Free 800 736 4666 Fax 724 334 5001 Website www mybacharach com Email help mybacharach com Printed in USA Registered Trademark of...

Reviews: