background image

Pur•Chek

 Pro

Instruction 2100-9001

11

If the refrigerant tested is 90% pure or better, the results will be displayed as shown in Figure 9. Should the 
refrigerant be less than 90% pure the results will be displayed as shown in Figure 10. 

NOTE: 

 Air is only displayed for R134A and only when the R134A refrigerant concentration is 

greater than  90%.  “NON” represents non-condensable such as air or another diluting gas.

4.7  Contaminated Blend Refrigerants

The Pur•Chek™ Pro includes the ‘Blend Chek Pro’ feature for determining the presence of R410A.  In the event 
that the Pur•Chek™ Pro determines that the primary refrigerant in the system or cylinder is likely to be R410A, 
the results will display as follows in Figures 10.  Pressing the “MORE” button will display the detail screen 
shown in Figure 11.  If the blend or refrigerant mixture is not recognized, the screen shown in Figure 12 will be 
displayed.

BACK

    MORE

Figure 10.

Figure 12.

Figure 11.

RESULTS

R410A

CONTAMINATED

RESULTS

R410A

98.7

R22

1.3

RESULTS

UNKNOWN

REFRIGERANT

BACK    PRINT    DONE

BACK    PRINT    DONE

4.8  Pure Blend Refrigerants

The Pur•Chek™ Pro has the ability to identify the following refrigerants: R408A, R409A, R417A, R421A, 
R421B, R422A, R422B, R422C and R427A.  In addition, the instrument can both identify and analyze the com-
ponent content of pure R134A, R22, R404A, R407C and R410A.  Analysis of analyzed pure blend refrigerants 
will yield additional data regarding the composition of the refrigerant sample. The Pur•Chek™ Pro will display 
the blend refrigerant type as shown in Figure 13.  Pressing the “MORE” button will display the detail screen 
shown in Figure 14. 

BACK

    MORE

Figure 13.

Figure 14.

RESULTS

R410A

RESULTS

R32

49.5

R125

50.5

BACK    PRINT    DONE

Summary of Contents for PUR-CHECK PRO

Page 1: ...Product Leadership Training Service Reliability Instruction 2100 9001 Operation and Maintenance Rev 0 June 2010...

Page 2: ...LIEU OF I ANY AND ALL OTHER WARRANTIES EXPRESS OR IMPLIED INCLUDING WITHOUT LIMITATION THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE AND II ANY OBLI GATION LIABILITY...

Page 3: ...3 6 Back Panel Connections 7 3 7 Hard Shell Storage Carrying Case 7 4 Operation 8 4 1 Battery Installation 8 4 2 Turning On the Unit 9 4 3 Calibration 9 4 4 Vapor Sampling 10 4 5 Liquid Sampling 10 4...

Page 4: ...Pur Chek Pro Instruction 2100 9001 ii...

Page 5: ...rformed or adhered to could result in minor or moderate injury IMPORTANT The use of the word IMPORTANT in this manual calls attention to a procedure practice condition or the like which if not correct...

Page 6: ...tain or will not emit heavy loads of oil or liquid NEVER admit any sample into the instrument at pressures in excess of 500 psig NEVER obstruct the air intake sample exhaust or case vent ports of the...

Page 7: ...The instru ment utilizes non dispersive infrared NDIR technology to determine the weight concentrations of multiple refrigerant types Refrigerant purity is displayed on the LCD Screen The user must de...

Page 8: ...rared NDIR Refrigerant Sample Size 0 3 ounces 8 5 grams per sample Power External Power Supply 110 220V 5 60Hz 0 6A External Battery Charger 110 220V 50 60Hz 1 95A DC Supply Requirements 12 15 VDC 2 A...

Page 9: ...void the warranty Printer Door Air Intake Port Sample Filter Control Panel 3 2 Low Side Vapor Sample Hose The 6 foot 1 8 meter Low Side Vapor Sample Hose is constructed of a polyurethane tube A brass...

Page 10: ...AE Flare Nut 3 4 AC Power Adapter The Pur Chek Pro is powered via a 110 220 VAC 50 60 Hz power transformer This transformer is included with each unit and converts a standard 110 220 VAC 50 60 Hz wall...

Page 11: ...let port free and clear at all times Do not operate near open flame Sample Inlet Sample Outlet Battery Charge Port 12 VDC Power Input AC Adapter 3 7 Hard Shell Storage Carrying Case The hard shell sto...

Page 12: ...rant Type R 12 R 134a R Miscellaneous Refrigerant Equipment 12 VDC 1A Design Pressure 300 psig CFC 12 R 12 or HFC 134a R 134a R 134a For Accuracy see Manual detect 2 or greater of air in HFC R134a to...

Page 13: ...alibration takes 95 seconds Figure 4 and brings fresh air into the unit via an internal pump This fresh air purges any excess refrigerant from the unit and ensures accurate test results Calibration re...

Page 14: ...om pletely depressed 3 Connect the appropriate end of the hose to the instrument and the opposite end of the liquid sampling assembly to the High Side Liquid port of the system or cylinder Affix the o...

Page 15: ...lay the detail screen shown in Figure 11 If the blend or refrigerant mixture is not recognized the screen shown in Figure 12 will be displayed BACK MORE Figure 10 Figure 12 Figure 11 RESULTS R410A CON...

Page 16: ...procedure To exit the test press the EXIT button Figures 15 16 17 show sample printouts for various test results NOTE Care must be taken to leave a clean edge when tearing off the printed results Neu...

Page 17: ...he ELEV button will take you to the elevation screen shown in Figure 19 Use the UP DOWN buttons to adjust the unit to the elevation in your area Press the SAVE button when finished to return to the he...

Page 18: ...Oil contamination cannot be cleaned out of sample hoses due to the density of the oil restrictor If oil is visible in the Low Pressure Vapor hose replace the hose and restrictor with P N 2100 0010 Fo...

Page 19: ...lled properly Slide the paper through the slot in the printer door and close the door 5 6 Error Messages In the unlikely event that an Error message is displayed on the screen power off the unit take...

Page 20: ...2100 0010 Spare Low Pressure Sample Hose Filter 2100 0012 Spare Rechargeable Battery 2100 9001 Instruction Manual 7 Service Centers Service and replacement parts can be found by contacting any of the...

Page 21: ...Pur Chek Pro Instruction 2100 9001 17 NOTES...

Page 22: ...ers 621 Hunt Valley Circle New Kensington PA 15068 PH 724 334 5000 Toll Free 800 736 4666 Fax 724 334 5001 Website www mybacharach com Email help mybacharach com Printed in USA Registered Trademark of...

Reviews: