PCA
3 Manual
60
0024-9473 Rev 3
5.13.4. Clearing Memory
When all memory locations used to store individual combustion test records have
been filled, the next combustion test record saved will not overwrite the oldest.
When all logging memory locations in the logging directory are full, they must be
manually cleared in order to store new data. At no time will the logging process
overwrite older data. Logging will stop when available memory is filled. Each
memory bank can be individually cleared, or all memory locations in both banks
can be cleared simultaneously. Do the following to clear memory.
1. Display the MAIN MENU by pressing the
MENU (F2)
button. If
necessary, press
ESC
until MENU appears above
F2
.
2. Use the ▲▼ buttons to highlight MEMORY, and then press
ENT
to
display the MEMORY MENU.
Figure 5-21. Clearing Memory
3. Use the ▲▼ buttons to highlight one of the following:
•
CLEAR MEMORY DATA clears only the individual saved
combustion test records.
•
CLEAR LOGGING DATA clears only the combustion test records
that were saved during the logging process.
•
CLEAR ALL MEMORY clears all memory locations in both memory
banks.
4. Press
ENT
to display the CLEAR MEMORY, LOGGING, or ALL DATA
conformation screen. Highlight YES to confirm that memory is to be
cleared, and then press
ENT
to actually clear memory as evidenced by
the display of the “WAIT Erasing memory” screen.
5.13.5. Print Average
The PCA
®
3 will automatically calculate, display, and print the average of three
consecutively saved combustion records. Calculate, display or print the average
data as follows.
1. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to highlight MEMORY, and then press ENT to
display the MEMORY MENU.
Summary of Contents for PCA3
Page 23: ...PCA 3 Manual 0024 9473 Rev 3 23 Figure 3 2 Connecting the Probe to the Analyzer...
Page 64: ...PCA 3 Manual 64 0024 9473 Rev 3 Figure 5 23 Printer Alignment and Sample Printout...
Page 84: ...PCA 3 Manual 84 0024 9473 Rev 3 Figure 7 2 Inside View with Rear Case Removed...
Page 86: ...PCA 3 Manual 86 0024 9473 Rev 3 Figure 7 5 Wire Routing Connectors and Solder Connections...
Page 101: ...PCA 3 Manual 0024 9473 Rev 3 101...
Page 102: ...PCA 3 Manual 102 0024 9473 Rev 3...
Page 103: ...PCA 3 Manual 0024 9473 Rev 3 103...