MNA0010, Revision B October 2014
5
Figure 1 Model 7100 Specifications
January 2001, Rev. 1
Sensor Type:
Mini-Zirconia Oxygen Sensor with Integral Heater.
Measurement
0-10ppm /0-100ppm / 0-2000ppm oxygen. May be set to auto range or to a fixed range.
Display:
0.75" LED digital display.
Resolution: 0-9.9ppm : X.X, otherwise XX
Color Coded LEDs for system status:
Alarms 1 & 2 =
Red
, System Fault =
Yellow
, System OK & on-line =
Green
Signal Interface
Serial Service Port: RS-232 Computer Communications
of the following data:
O2 concentration,
Oxygen Concentration, alarm settings/status, Fault status
Analog Outputs:
Voltage, 0-1 VDC or 0-5VDC or 0-10VDC +/- 0.5% @ 1K input min.
Non-isolated 4-20 mA: +/- 0.5% , negative ground (loop is powered by analyzer)
maximum loop impedance 500
. Range ID Voltage output
Alarm Outputs, optional
Relay dry contacts rated 5Amp @ 250 VAC, 5 AMP @ 30 VDC
Alarm 1 & Alarm 2: Field adjustable Form C,
Configurable for Fail Safe or Fail Alarm mode, Ascending or Descending trip.
Fault alarm: 1 Form B . Status LEDs for each alarm.
Response Time:
< 15 seconds for 90% response due to step change in ppm oxygen @ 1 SCFH flow;
approximately 5 minutes to less than 10 ppm from cold start after exposure to air
Accuracy:
+ 2 PPM at 0-50 PPM Oxygen
+ 3 PPM at 51- 500 PPM Oxygen
+ 10 PPM at 501- 2000 PPM Oxygen
* Enhanced accuracy when two point calibration at 500-700 ppm O2 & at 5-10 ppm O2
Warm up Time:
Approximately 5 minutes to reach thermal equilibrium with ambient temperature.
Power:
90-250 VAC,50/60 Hz., Single Phase USA or Euro Style
Sample Port:
1/8 inch Swagelok with SS sample tube for sample inlet
Operating Temperature:
5° to 40° C (31 - 104° F)
Storage Temperature:
-15° to 50° C (5 - 122° F)
Humidity:
0-95% non-condensing
Warranty:
Analyzer & sensor: 12 months from date of shipment under normal operation.
Operating Flow
Pressure
Sensor Flow: 0.7 to 1.2 SCFH; Total Flow: 1 SLPM
Pressure Range: 5” Hg. Column Gage Vacuum to 5 PSIG Positive Pressure
Sample Pump:
Neoprene diaphragm and valves, brushless vdc motor, plastic head
Weight:.
Less than 6 Lbs.
Mechanical:
9.10 inches wide x 7.27 inches deep x 7.15 inches high; less than 6 lbs. weight
Neutronics Inc., whose policy is one of continuous improvements, reserves the right to change specifications and this