MNA0010, Revision B October 2014
49
Table C3: Operational Table
Range Name
Measured Range
Display
Analog Range
Range Voltage Output
Auto Range
0-10 ppm
10-100 ppm
100-1000 ppm
0-2000 ppm
X.X
XX.
XXX.
XXXX.
0…10PPM
0… 100 PPM
0 …1000 PPM
0 …..2000 PPM
3.13 Volts 3.75
4.38
Fixed Range: 10PPM
0-10PPM
X.X
0…10PPM
3.13
Fixed Range:
100PPM
0-100PPM
XX.X
0…100PPM
3.75
Fixed Range:
2000PPM
0-2000PPM
XXX.X
0…2000PPM
4.38
Procedure to access settings at the 7100P Portable analyzer keypad:
1
To configure the 7100P Portable analyzer measurement range(s) through the front panel keypad, you must remove
all four of the hardware jumpers.
2
To enter the User configuration via the 7100P Portable analyzer keypad, Press and hold the MODE button for 10
seconds
. The display will show “---“. Release the mode button. You are now in the setup mode,
range setting
.
3
The display will show “A 1”. Pressing the mode key will scroll through the User configuration Modes.
MODE
INITIAL DISPLAY READING WHEN
SCROLLING
USER CONFIGURATION A: SET RANGE
A 0
USER CONFIGURATION 1: ALARM 1 TRIP
1 0
USER CONFIGURATION 2: ALARM 2 TRIP
2 0
USER CONFIGURATION 3: ANALOG OUTPUT
3 0
USER CONFIGURATION 4: SET SERIAL OUTPUT
4 0
USER CONFIGURATION 7: SET ZERO RANGE
7 0
USER CONFIGURATION 8: RESTORE FACTORY SETTINGS
8 0
USER CONFIGURATION E: SET VOLTAGE TO 110 OR 220 vac
E 0
USER CONFIGURATION F: SET FAIL SAFE
F 0
USER CONFIGURATION b: SERIAL BAUD RATE
b 0
FPT: FRONT PANEL TEST
“ – “ , then “ - - - “ after 10 seconds