background image

8A.  ZERO ADJUSTMENT

7:  OPERATION OVERVIEW

LED Operation

SHIELD WIRE WARNING:

   Connect the shield of 

Modbus wires to the earth ground of the central 

control system (e.g., chassis, ground bus bar, etc.).

Normal mode,

no alarm

Low gas alarm

High gas alarm

Fault

Under range 

(perform

zero adjust)

Over range 

(remove gas)

Offline (not in 

calibration 

mode)

Normal/Alarm/Fault/

Special States

Waiting to start 

calibration mode

During calibration

Zero calibration 

error

Zero Cal Mode

Waiting to start 

calibration mode

During calibration

Span calibration 

error

Span Cal Mode

Magnetic Switch Functions

Buzzer Operation

No alarm

Low gas alarm

Muted alarm

High gas alarm

Fault (continuous)

8:  GENERAL CALIBRATION PROCEDURE

1.

  The detector must not be in alarm or fault condition.  (One

exception is a negative gas concentration fault -- which requires

a zero calibration.)

2.

  Calibration gas must be in balance of air, 

not

 Nitrogen (N

2

).

3.

  Attach the pressure regulator to the calibration gas cylinder.

4.  Fit calibration adapter over sensor module, aligning its 3 

notches with the sensor alignment ribs.

5.  The gas flow should be approximately 0.3 to 1.0 L/min.

6.

  If operation is intended to be at higher altitudes, the factory 

calibration will result in a reading lower than the reading at sea 

level (reduced partial pressure).  A new span adjustment is 

recommended if the altitude or the ambient pressure is 

changed. The factory calibration is set to sea level.

7.  Connect the tubing to the barbed fitting.
8.  Always perform a zero adjustment before a span adjustment.

8B.  SPAN ADJUSTMENT

WARNING:  

Ambient air can be used to zero the 

sensor instead of synthetic air only if the area is 

known to be free of the target gas or any gas to which 

the sensor may be cross-sensitive. In this case, no 

cylinder or calibration adapter is needed for the zero 

adjustment.

9.    Tap and hold ( • ) for >5 seconds.  The LED will blink

 

green-green-red when the detector is ready.

10.    Apply synthetic air (or use ambient air per warning above).
11.   Tap ( • ) within 30 seconds to confirm start of calibration.

 

Otherwise the detector will time-out and return to normal

 

operation.

12.  As the process progresses, the LED will blink green-red,

 

green-red-red, green-red-red-red, etc.  

 

•  To abort calibration, tap and hold ( • ) for >5 seconds,

 

  turn off gas flow and remove the calibration adapter.

 

  The detector will return to normal operation.

 

•  If calibration is successful (green LED), skip to step 13.

 

•  If calibration is unsuccessful (orange LED blinks @ 2 Hz),

 

  then tap ( • ) to discard the calibration attempt, and see

 

  User Manual (P/N 6203-9000) for troubleshooting.

13.  Turn off gas flow from synthetic air.
14.  Replace synthetic air tank with calibration gas tank in

 

preparation for span adjustment (if required).

9.  BUMP TEST

Negative gas 

fault (perform

zero adjust)

Recovery from

span calibration

2.0 = 2.0 Hz

2 cycles/second

FAST flash/chirp

0.5 = 0.5 Hz

0.5 cycles/second

SLOW flash/chirp

LED Colors

6:  INSTALLATION (CONTINUED)

Warmup mode 

(≈ 6 minutes)

15. 

Tap and hold 

( •• ) for >5 seconds.  The LED will blink

 

green-green-orange when the detector is ready.

16.  Apply span gas in the concentration listed on the cal gas 

 

concentration label (beneath the detector’s cover plate). 

 

This may require the temporary removal of the bezel and

 

cover plate to see the label.

17.  Tap ( •• )

 within 30 seconds to confirm initiation of

 

the calibration.  Otherwise the detector will time-out and

 

return to normal operation.

18.  As the calibration process progresses, the LED will blink

 

green-orange, green-orange-orange,

 

green-orange-orange-orange, etc. 

 

•  To abort calibration, tap and hold ( •• ) for >5 seconds

,

 

  turn off gas flow and remove the calibration adapter.

 

  The detector will return to normal operation after

 

  6 minutes of recovery time.

 

•  If calibration is successful, the LED will blink

 

  green-orange-red indicating ‘offline’).   Turn off gas flow  

 

  and remove the calibration adapter.  Afte

r 6 minutes

  

the detector will return to normal operation.

 

•  If calibration is unsuccessful (orange LED blinks @ 2 Hz),

 

  then tap ( •• ) to discard the calibration attempt, and

 

  see User Manual (P/N 6203-9000) for troubleshooting.

 

  Turn off gas flow and remove the calibration adapter.

  

Afte

r 6 minutes the detector will return to normal 

 

 operation.

1. 

Inform building personnel of test so that certain alarms

 

may be inhibited (e.g., shutdown valves, notification of 

 

authorities, etc.).

2. 

Connect adapter and target gas according to instructions in

 

General Calibration Procedure.

3. 

Apply a sufficiently high concentration of target gas to

 

trigger alarms, but not pure refrigerant or hydrocarbons

 

(e.g., do not use a butane lighter).

4. 

Once the alarm thresholds are exceeded, all designated

 

gas alarm relays will be activated and the digital outputs 

 

will transmit the corresponding gas concentrations.

5. 

Turn off gas flow and remove calibration adapter.

High Gas Alarm

or Fault

Rela

y #2

NC

NO

COM

Low Gas Alarm

Rela

y #1

NC

NO

COM

Example of custom faceplate

Grounding lugs

Example of UK back box

Mounting tabs

Direction arrows (×2) for 

proper mounting

Sensor type/calibration and 

ID/serial number labels

Test point access holes (×2)

Replaceable sensor module

Sensor alignment ribs (×3)

5 s

8A

5 s

8B

5 s

5 s

5 s

Latched

alarm

60 min

60 min

Clears one

or both 

latched

alarms

Flow meter

Pressure regulator

Calibration gas

Tubing

Calibration adapter

Back view of a customized

faceplate (not included)

Sample

UK back box

(not included)

NOTE:

  For specifications and recommended drill 

patterns for customizing a metal UK faceplate, visit the 

MVR-300 product page at www.MyBacharach.com.

NOTE:

  For UK back boxes with “higher” mounting 

tabs,  remove the thick plastic mounting tabs from 

the detector (using wire snips or nippers) and screw 

on the thinner metal mounting tabs (included) to 

ensure flush wall mounting of the faceplate.  Be sure 

to dry fit the plastic tabs before cutting them from 

the detector to verfiy that they are too thick for 

proper flush mounting of the faceplate to the wall.

COVER GROUNDING:

   For proper cover ground-

ing, wire the ground lug of the faceplate to the 

ground lug of the earth grounded back box.

If the cus

tom f

acepla

te drill pa

ttern does not acc

ommoda

te  the tw

o magne

tic s

wit

ch

loc

ations, the f

acepla

te mus

t be r

emo

ved t

o use the magne

tic w

and functions.

Reviews: