When finished adjusting the settings, press ‘Exit’ to return to the home screen.
3.5 Error Messages
In the unlikely event an Error message is displayed on the screen, follow the on-screen
prompt associated to the Error. Error messages that will appear include:
Error #1: The air or gas readings were unstable.
•
Solution: Rerun test
•
Solution: If problem persists contact manufacturer
Error #2: The air or gas readings were excessively high.
•
Solution: Rerun test
•
Solution: If problem persists contact manufacturer
Error #3: The air calibration resulted in a low sensor output.
•
Solution: Prevent refrigerant from flowing into the unit through the sample inlet during
air calibration.
•
Solution: Verify no oil contamination is present in sample hose or sample filter
•
Solution: Verify that the air intake and the exhaust are not obstructed
•
Solution: Verify that the white filter is correctly plugged into the rubber grommets.
•
Solution: Allow any refrigerant in the atmosphere to dissipate before performing air
Error #4: The unit is beyond the operating temperature range.
•
Solution: Move the unit to an area where the ambient temperature is within the
specified operating range.
Error #5: The refrigerant sampled has an excessively large amount of air or there was a little or no
sample flow due to a closed valve or plugged sample filter. This is the code to prompt the user to
change the brass filter. This should be considered more as a prompt than an actual error.
•
Solution: Verify the refrigerant source is open and flowing.
•
Solution: Verify the sample filter is not plugged with debris or oil.
•
Solution: Replace brass sample filter.
Error #6: The air sensor has expired and must be replaced before the analyzer can be used.
•
Solution: If problem persists contact manufacturer
Error #7: The gas pressure is out of range
•
Solution: Verify the SAMPLE EXHAUST port is not obstructed.
If an Error message reappears, contact Bacharach or your local service department.