Manual Part Number:
5-06-7000-81-1, 7-
08-1234-40-0
Manual File:
069706 Rev. A
Page 2-16
Once the analyzer warms up, the screen in
(Figure 3)
will appear offering the option to
change settings or start an analysis. If you wish to adjust factory ‘Settings’, select the left soft
key and refer to section
3
Maintenance & Troubleshooting.
To begin an analysis, select the right soft key to ‘Start.’ Next use the left and right soft keys to
scroll through available refrigerant options and the center soft key to select the refrigerant
you wish to test
(Figure 4).
2.3 Calibration
Each time the
Legend Series™
HFC begins a new test cycle it must complete an air
calibration. The calibration takes 30 seconds and pulls fresh air into the unit via an internal
pump. This fresh air purges any excess refrigerant from the unit and ensures accurate test
results. Calibration
REQUIRES
the ¼” Flare Vapor sample hose be connected to the device
and disconnected from the refrigerant source.
Once the sample hose is connected to the analyzer, press ‘Start’ to begin an air calibration,
as shown in
(Figure 5)
. This will begin the calibration process and display the Calibrating
screen shown in
(Figure 6
).
Figure 3
Figure 6
Figure 5
Figure 4