Fyrite INSIGHT
51
- Range: 0 to 300 °C
- Accuracy: ± 0.3 °C
•
(Alternatively) ice-water, boiling water, thermometer
TS-Zero Procedure:
1. Set the thermocouple simulator to room temperature and plug its
output into the T-STACK connector located at the bottom of the
analyzer.
Alternatively:
Plug the probe’s thermocouple into the T-STACK con-
nector located at the bottom of the analyzer.
DO NOT attach the
probe’s gas hose to the analyzer’s GAS port; otherwise water
will be drawn into the analyzer!
2. If not already done, power up the analyzer and display the CALI-
BRATION LIST screen per Section 5.2.
3. Use the ▲▼ buttons to highlight T-Stack, and then press
ENTER
to
display the CALIBRATE TS-ZERO screen.
1
“Measured” is the current temperature reading, while “Applied” is a
known temperature that will be applied for calibration purposes.
4. Set thermocouple simulator to 0 °C (32 °F), and then use the
and
buttons to enter an Applied value that exactly equals the setting
of the simulator.
Alternatively:
Submerge probe tip into an ice-water bath with a ther-
mometer, wait several minutes, and then use the and buttons
to enter an Applied value that exactly equals the thermometer read-
ing.
The calibration range is from 0 to 5 °C (32 to 41 °F ). An attempt to
calibrate outside this range will cause the message “Applied Value
High” (or Low) to appear at the bottom of the screen.
5. Wait until the Measured reading stabilizes, and then press
ENTER
to calibrate the TS-Zero Measured value to that of the Applied
Summary of Contents for 0024-8252