See illustration 8 for layout of control unit featuring:

• On/Off button.

• Green movement indicator light.

• Acoustic and visual red alarm light.

• Low battery red warning light.

• Yellow shutdown indicator light

Babysense

 is ready for operation and use once the 

installation is completed. To operate, follow the below 

instructions:

• Place your baby in crib/bed.

• Turn control unit on by pressing On / Off button.

• Make sure red control lights switch on briefly and a 

DOUBLE click sound is heard. 

• When any movement is sensed, the green   

light flashes. 

• Acoustic and visual red alarm light are activated if 

the baby's movement stops or slows significantly.

• If red low-battery warning light flashes, replace all 

batteries as soon as possible. If in addition clicking is 

heard, all batteries must be replaced immediately. 

• Turn off by pressing On / Off button before taking 

your baby out of crib / bed. Make sure the Yellow 

shutdown indicator light momentarily blinks and no 

other lights are seen thereafter to indicate the device 

is turned off.

Due to the high sensitivity of Babysense, fans, wind, air 

conditioning, and so forth can influence it. Such 

environmental effects could be picked up by the sensors 

on a similar level to that of slight movements, thus 

preventing or delaying alarm activation.

If the Babysense is concurrently used with other devices, 

it is required to conduct the daily performance test 

(below) under an environment identical to the actual 

situation and while all concurrently used devices are 

"ON". Under such conditions, it is required to verify that 

the acoustic and visual alarms of Babysense are activated. 

 

Do not conduct performance test by picking up baby 

out of bed and waiting for the alarm to sound   

• The performance test should be conducted while baby 

is not in bed. 

• Turn device on and take two steps away from bed. 

• Ensure that alarm goes off within 20 seconds (due to 

absence of movement).

• The alarm may sound with or without preliminary ticking. 

• Check that alarm can be heard in all locations 

throughout the home or place of care.

• In case that alarm does not sound within 20 seconds 

refer to troubleshooting on page 4-5.

• Performance test must be conducted daily and every 

time you move the monitor, change bedding or 

relocate crib/bed position.

 

Isolated cases of false alarms are possible. Check if 

caused by faulty connection of the sensor pad(s) to the 

control unit, connection of only one sensor pad when 

baby can move off the immediate area or failure of the 

sensor pads to be directly against the mattress.

In case of false alarms, kindly refer to troubleshooting 

guide on page 4-5. If the problem cannot be solved, 

please contact manufacturer’s service personnel.

 

Layout of control unit

Operation

Environmentally caused malfunction

Conduct this performance test daily

False alarms

I

llustration 8

Green movement 

movement 

indicator light

Red

alarm light

Red low battery 

warning light

Hidden On/O

ff

power button

3

User Guide

Yellow shutdown

indicator light

8. 

Press control unit holder or hanging hook firmly into 

slot in back of control unit (see illustration 4 and 5).

9. 

Place mattress firmly on top of sensor pads. If the 

mattress cannot be in contact with the sensor pads, or if 

the bed base is soft, padded or contains springs, place 

extra plywood on the bed base, place the sensor pads on 

the plywood and place the mattress firmly on top.

IMPORTANT: 

To avoid possible danger from sensor 

cords, ensure that excess cords are pulled back under 

the mattress, secured with the included cable tie and 

are not visible or reachable.

 

Testing Operation

Technical malfunction

Babysense 

is a highly sensitive device, which monitors a 

baby's motion and movements through the mattress 

during sleep. There are two main causes of malfunction - 

technical malfunctions and malfunctions due to 

environmental factors.

 

In order to minimize technical problems, follow 

installation and operation instructions carefully.

Summary of Contents for 7

Page 1: ...User Guide English Since 1991 7 The original non contact baby movement monitor Now with enhanced sensitivity ...

Page 2: ...purpose of the device is to alert when stopping or slowing of movement occurs Babysense is a unique safety and reassurance system which helps give parents peace of mind alerting if movements stop or become irregularly slow giving critical time to intervene There are a number of cautionary steps which are internationally recommended for SAFE SLEEPING Babysense continually detects baby s motion and ...

Page 3: ...not use Sensor Slot Blocked socket DO NOT USE Sensor socket Battery cover Opening recess Speaker Illustration 7 Remove batteries when not in use Sensor Do not use Sensor Slot Illustration 6 Remove batteries when not in use Sensor Do not use Sensor Slot Optional Hanging hook Plywood Illustration 3 Sensor Sensor Spring base Top crib bed view User Guide 6 Attach end s of the cord s to the outlets mar...

Page 4: ... 20 seconds due to absence of movement The alarm may sound with or without preliminary ticking Check that alarm can be heard in all locations throughout the home or place of care In case that alarm does not sound within 20 seconds refer to troubleshooting on page 4 5 Performance test must be conducted daily and every time you move the monitor change bedding or relocate crib bed position Isolated c...

Page 5: ...oltage warning red light on the Control Unit False alarm up to 1 2 minute from activating False alarm after 1 2 minute from activating Delayed alarms while taking the baby out of the bed Environmental influences detected by the sensor pads such as wind fan air conditioner baby mobile toy vibrations Perform a test which will detect the source of the movement that the device detects For example sourc...

Page 6: ...acturer for recycling or proper disposal Babysense requires no special treatment beyond replacing batteries when the battery indicator light goes on The sensor pads and control unit may be wiped with a damp cloth Batteries should be removed and the device stored in a dry shaded place when not in use Caring for your babysense It is recommended to replace sensor pads two years after their first use U...