47
EMC compatibility
The Babyation breast pump requires special precautions regarding EMC and needs to be installed and
put into service according to the EMC information provided in the instructions for use. Portable and
mobile RF communications can affect the breast pump.
This breast pump is intended for use in the electromagnetic environment specified below. The customer
or use of the breast pump should assure that it is used in such environment.
Electromagnetic emissions
Emission tests
Compliance
Electromagnetic Environment Guidance
RF Emissions
CISPR 11
Group 1
The Babyation breast pump uses RF energy
only for its internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in the nearby
electronic equipment.
RF Emissions
CISPR 11
Class B
Harmonic emissions
IEC-61000-3-2
Class A
Voltage fluctuations/flicker
EN 61000-3-3
Complies
Guidance and manufacturer’s declaration – immunity
Immunity test
IEC 60601 test level Compliance level
Electromagnetic
Environment Guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
That relative humidity
should be at least 5%
Electrical fast
transient/burst
IEC 61000-4-4
Mains power quality should
be that of a typical home,
commercial, or hospital
environment.
Surge
IEC 6100-4-5
Mains power quality should
be that of a typical home,
commercial, or hospital
environment
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
Mains power quality should
be that of a typical home,
commercial, or hospital
environment
If the user of the Babyation
breast pump system
requires continued
operation during power