background image

Alarm

If the baby is not breathing for longer than 12 seconds, or the breath rate is too slow, after 4 
warning signal an alarm signal will sound and the red indicator will flash. On the display(2) 
the „H”-„E”-„L”-„P” letters will be displayed. (If movement is detected during the 4 warning 
signal the alarm will not sound.
Check the baby and provide first aid if necessary (make sure that there are no reasons for 
choking, start resuscitation, call for emergency help). 
In some cases of Apnea the sound of the unit itself can wake the baby up and he/she will start 
to breath again on his/her own. 
Alarm can be stopped by the switching the unit off.
There may be rare cases of false alarms, especially when the child is crawling and moves off 
the sensor pad. 

E

   vent Memory

 

 

The control unit is capable to store the event of the last 10 minutes before the alarm.
In case of an alarm it is vital to know and understand the previous events for an examination 
process.

Review

 

  the event memory:

 

 

1. Turn off the unit and wait a few seconds

2.

Then push the red button(1) and hold it for 10 seconds until you see the  „P” and „L” 
letters on the display

3. Release the button 

4.

On the display you can select which minute you would like to see ,„1 P”, „2 P” ... „9 P” 
(changes slowly in the display).

5. When you see the required time press the button
6. The measured values will begin to show up

Press the button again and the unit will return to the normal mode.

Battery 

 

 Replacement

 

 

If the batteries are low, the red indicator will flash and the digital display shows error "E", 
there will be beep sound after it is tuned on.  
For safety reason the unit can not be switched off until you remove/replace the batteries.
In this case replace the batteries. Use only 2 new size AA alkaline batteries (orientation is 
clearly marked in the battery compartment). Switch the BABY CONTROL on after battery 
replacement and you should hear a beep sound. If there is no beep, check the batteries. 

Important

: Do not use rechargeable accumulators!

Summary of Contents for BC-200

Page 1: ...r pad is 30 x 50 cm Depending on the product having 1 2 3 or 4 twins sensor pads The product with 3 sensor pads is able to cover the whole resting place of the baby even large cots Extra Sensor Pads E...

Page 2: ...sess these features in the market Product Packages A BC 200 package contains 1 BC 103 control unit 1 BC 101 sensor pads A BC 210 package contains 1 BC 103 control unit 2 BC 101 sensor pads A BC 230 pa...

Page 3: ...So it cannot have any harmful effect to the human body IMPORTANT Do notice that BABY CONTROL can only alert you when your child is breathing irregularly or not at all It cannot make any action itself...

Page 4: ...p after turning ON check the batteries 3 Attaching the control unit to the bed using the fastening belt The unit should be visible and should not be covered by anything For bigger children we recommen...

Page 5: ...e digital display 2 then release the button 3 On the display the numbers are changing slowly from 1 to 9 4 When the required sensitivity threshold level is displayed press the red button 1 5 Now the n...

Page 6: ...understand the previous events for an examination process Review the event memory 1 Turn off the unit and wait a few seconds 2 Then push the red button 1 and hold it for 10 seconds until you see the...

Page 7: ...inute or so if not the device is detecting some intrusive vibrations see the following warning Important warning The BABY CONTROL is a very sensitive instrument capable of sensing a sleeping baby s br...

Page 8: ...eration has to be switched off e g one of the baby is taken out from the bed 5 Then press the button 6 The device is switched to single mode 7 When you put the baby back the press the button again thi...

Page 9: ...a considerably expanded research effort However there are things that can be done to reduce the risk of SIDS Parent To Be 1 Get medical care early in pregnancy preferably within the first three month...

Page 10: ...ater the exposure to tobacco smoke the greater the risk of SIDS 7 Breast feed babies whenever possible Breast milk decreases the occurrence of respiratory and gastrointestinal infections Studies show...

Page 11: ...f Comformity Manufacturer VMD 95 Kereskedelmi s Szolg ltat Kft The subject of the declaration Baby Control Digital Apnoe Alarm Systems The manufacturers declares that the product to which this declara...

Page 12: ...al purchase If this system fails to function properly when used as directed under normal conditions within the two year period the Manufacturer will either repair or replace the product at our option...

Reviews: