2 After starting a
cleaning cycle,
isopropyl alcohol is
not being pumped
into the inner
container.
The outer container is inserted
into the base backwards.
Rotate the outer container so
its pump tube is near the front
of the
B9Clean
base.
The inner container is inserted
into the outer container
backwards.
Press the Start/Stop button to
stop the cleaning cycle and
wait for the isopropyl alcohol
to drain from the inner
container.
Rotate the inner container so
its pump tube opening is near
the front of the
B9Clean
base
and restart the cleaning cycle.
Check sealing at bottom side of
container. Black sealing
material can dissolve over
time.
Replace foam sealing ring at
bottom side of inner container
with original spare part. See
chapt. 10 – service.
If none of the above solves the
problem, contact B9Creations,
LLC, support.
3
During a cleaning
cycle, the isopropyl
alcohol in the inner
container is not
being agitated.
The magnetic stirring bar is not
placed in the bottom of the
inner container.
Press the Start/Stop button to
stop the cleaning cycle and
wait for the isopropyl alcohol
to drain from the inner
container.
Place the magnetic stirring bar
in the bottom of the inner
container and restart the
cleaning cycle.
If none of the above solves the
problem, contact B9Creations,
LLC, support.
4
During a cleaning
cycle, the isopropyl
alcohol in the inner
container is not
completely
submerging the
3D printed parts.
There is not enough isopropyl
alcohol loaded into the outer
container.
See section
7
to determine
how to set up the
B9Clean
based on the height of the 3D
printed parts being cleaned.
The inner container drain plug
is in the wrong position.
The height of 3D printed parts
exceeds the maximum allowed
height of 127 millimeters
(5.125”).
Refrain from using the
B9Clean
to clean parts that exceed the
maximum height allowance.
If none of the above solves the
Summary of Contents for B9Clean
Page 1: ...USER MANUAL B9Clean...