4 Design and Construction
24
DHV0075 b2 EN
©
b2 electronic
gmbh
Nr.
Name
Description
D
TD system control
By removing the aluminium head cap the main switch and the
battery housing can be accessed.
Main Switch
Switch for powering up or shutting down the TD system. It is
advisable to turn off the system to conserve battery life if the
system is not planned to be used for an extended period of
time.
The TD system incorporates an automatic power save sleep
mode, which shuts down the system after no voltage (<1kV)
has being sensed for 60 minutes or more. In this mode the
main switch has to be set to 0 (off) for about 2 minutes and
then on again to wake up the system manually, or you can
reapply voltage with the HVA which will also wake up the unit.
Battery / accumulator - Box
Just unscrew the lid of the no to change the battery /
accumulator
TD-MC
battery Size D
TD
battery Size C
HV output connector
Serves as connection point from the PD-2CC to the DUT via
HV test lead.
• Plug HV test lead into the HV output connector
Guard Connection
Connection for Guard (leakage current detection and
correction)
(Accessories for Guard Detection isn’t included in the
Standard Accessories)
D
TD60
TD60-MC