Memory Deletion

1. Remove one of the batteries, long press the ON Button.
2. Reload the battery until “ 

 ” on the LCD screen.

 

    

 

Automatically on the 26th measurement: when the 25 memories have 

been used up, any new measurement will be recorded and the oldest 
memory deleted without you having to do anything.

 

NOTE:

All the readings will be cleared no matter record in Body Mode or 

Object Mode.

15. CLEANING INSTRUCTIONS

For the device disinfection, 75% alcohol (available in the pharmacy) 
can be used. 

Measurement Sensor:

Please wipe with a clean and dry cotton swab to clean the Sensor on 
the inside. Do not use water, the Sensor is not waterproof.

Thermometer:

Use a soft and dry cloth to clean the thermometer body. Do not use 
abrasive cleaners. The body of the thermometer is not water-resistant. 
Never put the thermometer under a running tap or submerge it into 
water. 
Store the thermometer in a cool and dry location. Free from dust and 
away from direct sunlight. Do not expose the thermometer to extreme 
temperatures, humidity, or shock.

16. TROUBLESHOOTING 

FAULT OR 
FAULT SYMBOL

FAULT DESCRIPTION

CORRECTIIVE 
MEASURE

No display On 

the LCD pane.

The battery has run 

out. Incorrect battery 
polarity.

Replace the 
battery.
Please note:

The (+) side of 

the battery must 
face Positive (+) 
indication on the 
device.

Measurement 

not possible (or 

an abnormal 

value is 

displayed).

The thermometer is not 

ready.

Wait until 
the °C symbol is 
displayed.

An abnormal 

temperature 

value is 

displayed.

The Sensor is dirty or 

damaged.

Clean the Sensor 
or get it repaired.

The thermometer was 

removed too early 
(before the beep sound).

Wait until you 
hear the beep 
before removing 
the thermometer 
from the forehead.

or 

 

symbol is 

displayed.

The temperature 

measured is outside the 
measuring range.

Body Mode: 

LO

 – temperature 

<32°C (89.6°F).

HI 

 – temperature 

≥42,9°C (109.3°F)

Object Mode: 

LO

 – temperature 

<0°C (32°F).

HI

  – temperature 

≥100°C (212°F).  

Check that the 
probe tip is clean 
and that the 
thermometer is 
properly placed 
on the forehead.

 

 Symbol is 

displayed.

The battery has run out.

Replace the 
battery.

Err

Symbol is 
displayed.

The ambient 

temperature is 
outside the operating 
temperature range or is 
changing too fast.

To ensure accurate 

measurement, let 
the thermometer 
rest at operating 
temperature for 

30 minutes prior 

to use.

22. SYMBOL INFORMATION

IP20

FOLLOW INSTRUCTIONS BEFORE USE 

MANUFACTURER’S NAME

ARTICLE NUMBER

SERIAL NUMBER

TYPE BF EQUIPMENT

DISPOSAL FOR SEPARATE COLLECTION

CE MARK (0044)

OPERATING CONDITION, 
TEMPERATURE 5˚C ~ 40˚C

STORAGE CONDITION, 
TEPERATURE -25˚C ~ 70˚C

KEEP DRY

HOUSING INGRESS PROTECTION RATE: IP20

0044

40

  5

-25

70

17. APPLIED STANDARDS

Device Standards:

Device Corresponds to the requirements of the standard for infrared 
thermometers
IEC 60601-1:2012
IEC 60601-1-2:2014
ISO 80601-2-56:2009

Classifi cation:

Anti-electric Shock Type: Internally powered equipment
Applied part: Type BF
Mode of operation: Continuous Operation
EMC: type B class I

Electromagnetic Compatibility:

Device fulfi lls the stipulations of the standard IEC 60601-1-2

The stipulations of EU-Directive 93/42/EEC for Medical Devices 

Class Ila have been fulfi lled.

18. PRODUCT SPECIFICATIONS

Measurement range:

     Human Body:

     Object:

32,0 ~ 42,9°C (89,6°F ~ 109,3°F)

0,0°C ~ 100°C (32,0°F ~ 211,8°F)

Accuracy:

     For Body mode:

     For Object mode:

±0,2°C (±0,4°F): 
36,0°C ~ 39,0°C (96,8°F ~ 102,2°F)
±0,3°C (±0,5°F): out of the range

±1°C (±2°F): 

0°C ~ 60,0°C (32,0°F ~ 140,0°F)

±4°C (±7,2°F): out of the range

Display resolution: 

0,1°C or 0,1°F

Temperature unit:

Default °C or Default °F

Operation Distance:

3 cm

Traffi c lights backlight 

(Body mode), Indicator 
light range:

Green: <37,5°C (<99,5°F);

Yellow: ≥37,5°C - ≤37,9°C (≥99,5°F - 
≤100,2°F);

Red: 

>

37,9°C (

>

100,2°F)

Beeps defi nition:

Power on and ready for working: a 
short beep
Measurement fi nished:

1 long beep ≤37,9°C (100,2°F)
6 short beeps >37,9°C (100,2°F)

Memory:

25 mesurements 

Auto shut-off:

about 10 seconds after no using

Batteries: 

2 x 1.5V AAA size alkaline batteries

Dimensions (L×W×H): 

142,5 mm × 38 mm × 40 mm

Weight:

76g (Net: 54g)

Operating environment:

Temperature: 

+5°C ~ +40°C (41°F ~ 104°F);

Humidity: 

15% ~ 93 % RH (non-condensing);
70Kpa-106Kpa.

Transport and Storage 

environment:

Temperature: 

-25°C ~ +70°C (-13°F ~ 158°F);
Humidity: 
15% ~ 93% RH (non-condensing).

19. THERMOMETER SET

Thermometer set:

1. Thermometer
2. Battery 2 x 1.5V AAA size alkaline batteries
3. Soft bag
4. Instructions for  use

20. UTILIZATION

The unit must be utilized in accordance with current standards 

separately from domestic wastes. For utilizing it is necessary to 
contact special organizations licensed to make utilization.

21. WARRANTY

Warranty period is 2 years from the date of purchase. 

This warranty doesn’t cover any damages caused by improper 

using, and also battery, protective cover and packaging. When a 
manufacturing defect is revealed during the warranty period a faulty 
unit would be repaired or, if repairing is impossible, replaced with 
another one. 

Manufacturing date is on the unit in a serial number: last 2 fi gures of 
the year, then month number. 

The manufacturer may change units partially or completely if 

necessary, without prior notice.

ELECTROMAGNETIC COMPATIBILITY INFORMATION

Manufacturer’s Declaration of the Electromagnetic 
Compatibility

Table 1

Guidance and manufacturer’s declaration – 

еlectromagnetic emissions

Emissions test

Compliance

RF emissions
CISPR 11

Group 1

RF emissions 
CISPR 11

Class B

Harmonic emissions 
IEC 61000-3-2

Not application

Voltage fluctuations/ flicker emissions
IEC 61000-3-3

Not application

Table 2 

Guidance and manufacturer’s declaration – 

еlectromagnetic immunity

Immunity test

IEC 60601-1-2 
Test level 

Compliance level

Electrostatic discharge 
(ESD) 
IEC 61000-4-2 

±8 kV contact 
±2 kV, ±4 kV, ±8 kV, 
±15 kV air

±8 kV contact 
±2 kV, ±4 kV, ±8 kV, 
±15 kV air

Electrical fast transient/
burst 
IEC 61000-4-4 

Power supply lines: ±2 kV 
input/output lines: ±1 kV 

Not application 

Surge 
IEC 61000-4-5

line(s) to line(s): ±1 kV. 
line(s) to earth: ±2 kV. 
100 kHz repetition 
frequency

Not application 

Voltage dips, short 
interruptions and voltage 
variations on power supply 
input lines 
IEC 61000-4-11 

0% 0.5 cycle  
At 0°, 45 °, 90 °, 
135 °, 180 °, 225 °, 270 ° 
and 315 °  
0% 1 cycle  
And  
70% 25/30 cycles  
Single phase: at 0 
0% 300 cycle 

Not application 

Power frequency 
magnetic fi eld 
IEC 61000-4-8

30 A/m
50Hz/60Hz

30 A/m
50Hz/60Hz

Conduced RF 
IEC 61000-4-6

150KHz to 80MHz: 
3Vrms 
6Vrms (in ISM and 
amateur radio bands) 
80% Am at 1kHz

Not application

Radiated RF 
IEC 61000-4-3

10 V/m 
80 MHz – 2,7 GHz 
80 % AM at 1 kHz

10 V/m 
80 MHz – 2,7 GHz 
80 % AM at 1 kHz

NOTE  

U

T

 is the a.c. mians voltage prior to application of the test level. 

Remark In the presence of strong electromagnetic fi eld (Horizontal/Vertical, 
80MHz~180MHz and 1.4GHz~2GHz, 10V/m applied on the product), it will have 
a change from 0.1°C to 0.3°C of the measured value. In case of malfunction due 
to these strong electromagnetic fi elds, user need to wait till resume normal 
operation without intervene.

IM_MED-

3000_EN_5

120

 B.Well Swiss AG

Bahnhofstr. 24, 9443 Widnau, Switzerland

www.bwell-swiss.ch

Table 3

Guidance and manufacturer’s declaration – 

electromagnetic Immunity

Radiated RF 
IEC61000-4-3 
(Test 
specifi cations 
for ENCLOSURE 
PORT IMMUNITY 
to RF wireless 
communications 
equipment) 

Test

Frequency

(MHz)

Band

 

(MHz)

Service

M

odulation

M

odulation

(W)

Distanc

e

(m)

IMMUNITY

TEST 

LEVEL

(V/m)

385 

380 –
390

TETRA 400

Pulse

modulation

18 Hz

1,8

0.3

27

450 

380 –
390

GMRS 460,

FRS 460

FM

± 5 kHz

deviation

1 kHz sine

2

0.3

28

710

704 –
787

LTE Band 

13, 17

Pulse

modulation

217 Hz

0,2

0.3

9

745 

780

810

800 –
960

GSM

800/900,

TETRA 800,

iDEN 820,

CDMA 850,

LTE Band 5

Pulse

modulation

18 Hz

2

0.3

28

870

930

1720

1 700 –
1 990

GSM 1800;

CDMA 

1900;

GSM 1900;

DECT;

LTE Band 1, 

3, 4, 25; 

UMTS

Pulse

modulation

217 Hz

2

0.3

28

1845

1970

Last revision 2020-W51

Summary of Contents for MED-3000

Page 1: ...sing Always operate the thermometer in an operating temperature range 5 C to 40 C 41 F to 104 F and relative humidity 15 to 93 Always store the thermometer in a cool and dry place 25 C to 70 C 13 F to 158 F and relative humidity 15 to 93 The device requires no calibration The device contains no user serviceable parts The user must check that the equipment functions safely and see that it is in pro...

Page 2: ... in mind that temperatures measured from different sites even at the same time should not be directly compared Fever indicates that the body temperature is higher than normal This symptom may be caused by infection overdressing or immunization Some people may not experience fever even when they are ill These include but are not limited to infants younger than 3 months old persons with compromised ...

Page 3: ... used to measure the object temperature Choose the mode you need using a slider on the side of the device Silent mode Beeps may be switched off 1 Be sure the infrared thermometer is off 2 Hold the ON Button for about 12 seconds after ON shows on the LCD panel Note keep holding the button when C F is displayed on the LCD panel 3 Press the ON Button again to turn ON the beep 4 Then long press the Bu...

Page 4: ...l briefly display all of its symbols during this time 2 Push the Mode switch to select Object mode The device sounds two short beeps indicating that the object mode is turned on and the temperature unit flashes 3 Position the thermometer at a distance within 3 cm from the object Make sure the probe is flat and close to the object not at an angle 4 Press and release the ON Button The measurement resul...

Page 5: ...UBLESHOOTING FAULT OR FAULT SYMBOL FAULT DESCRIPTION CORRECTIIVE MEASURE No display On the LCD pane The battery has run out Incorrect battery polarity Replace the battery Please note The side of the battery must face Positive indication on the device Measurement not possible or an abnormal value is displayed The thermometer is not ready Wait until the C symbol is displayed An abnormal temperature ...

Page 6: ...nt finished 1 long beep 37 9 C 100 2 F 6 short beeps 37 9 C 100 2 F Memory 25 mesurements Auto shut off about 10 seconds after no using Batteries 2 x 1 5V AAA size alkaline batteries Dimensions L W H 142 5 mm 38 mm 40 mm Weight 76g Net 54g Operating environment Temperature 5 C 40 C 41 F 104 F Humidity 15 93 RH non condensing 70Kpa 106Kpa Transport and Storage environment Temperature 25 C 70 C 13 F ...

Page 7: ...ctrical fast transient burst IEC 61000 4 4 Power supply lines 2 kV input output lines 1 kV Not application Surge IEC 61000 4 5 line s to line s 1 kV line s to earth 2 kV 100 kHz repetition frequency Not application Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 0 0 5 cycle At 0 45 90 135 180 225 270 and 315 0 1 cycle And 70 25 30 cycles Single ph...

Page 8: ... Modulation Modulation W Distance m IMMUNITY TEST LEVEL V m 385 380 390 TETRA 400 Pulse modulation 18 Hz 1 8 0 3 27 450 380 390 GMRS 460 FRS 460 FM 5 kHz deviation 1 kHz sine 2 0 3 28 710 704 787 LTE Band 13 17 Pulse modulation 217 Hz 0 2 0 3 9 745 780 810 800 960 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modulation 18 Hz 2 0 3 28 870 930 1720 1 700 1 990 GSM 1800 CDMA 1900 GSM 1900...