background image

INTRODUCTION

This instrument is a portable easy to use 3 digit, compact-

sized digital infrared thermometer with laser sighting de-
signed for simple one hand operation. Meter comes with
Backlit LCD display, Auto-hold function and auto power
down (

15 seconds

 approx.) after releasing MEAS button to

extend battery life.

SAFETY  INFORMATION

It is recommended that you read the safety and operation

instructions before using the infrared thermometer.

CAUTION

• Do not use the unit near any device which generates

strong electromagnetic radiation or near a static

electrical charge, as these may cause errors.

• Do not use the unit where it may be exposed to

corrosive or explosive gases. The unit may be dam-

aged, or explosion may occur.

• Do not keep or use this unit in an environment where

it will be directly illuminated by sunshine, or where it

will be exposed to high temperatures, high humidity or

condensation. If you do, it may be deformed, its

insulation may be damaged, or it may no longer

function according to specification.

• Do not point the lens at the sun or at any other source

of strong light. If you  do, the sensor may be damaged.

• Do not contact the lens against the object whose

temperature is to be measured, or get it dirty, allow it

to be scratched, or allow any foreign material to

adhere to it. Doing so may cause errors.

• Do not touch  or hold by the front cone. Temperature

reading can be affected by heat from hand.

• Do not place the meter on or around hot objects (70°C/

158°F). It may cause damage to the case.

• If the meter is exposed to significant changes in

ambient temperature (hot to cold or cold to hot). Allow

20 minutes for temperature stabilization, before tak-

ing measurement.

• Condensation may form on the lens when going from

a cold to hot environment-wait 10 minutes for conden

sation to dissipate before taking measurements.

• This unit is not constructed to be water proof or

dustproof, so do not use it in a very dusty environment

or in one where it will get wet.

SPECIFICATIONS

GENERAL

Display:

3½ digit liquid crystal display (LCD) with maximum
reading of 1999

Low battery indication: 

 the "      " is displayed when the

battery voltage drops below the operating level

Measurement rate: 

1 times

  per second, nominal.

Operating Environment: 

32°F to 122°F (0°C to 50°C)

at < 80% relative humidity

Storage Temperature: 

40°F to 140°F (-20°C to 60°C)  ,

0 to 80% R.H. with battery removed from meter

Auto power off:

 

15 seconds

.

Standby consuming current:

 <

20

µ

A

Battery:

Standard 9V battery (NEDA 1604, IEC 6F22)

Battery Life:

 

100

 hours (continuity) typical with carbon-

zinc battery (Laser marker and Back-Light not illumi-
nated)

Dimensions:

170mm(H) x 65.5mm(W) x 35mm(D)

Weight:

 195g (including probe and batteries)

Laser Specifications:

Laser safety classification of Class 2

Wave Lenght:

 Red (630nm~670nm).

Operating Distance:

 2 to 50 feet.

Power out:

 <1mW , class 2 laser product.

 DANGER

Pressing the   button turns the laser beam on and off.
Exercise extreme care and do not allow the laser beam
to enter your eye or those of any other person or animal.
• Do not look directly into the laser light from the optical

system.

• When measuring the temperature of an object which

has a mirror finish, be careful not to allow the laser
light beam to be reflected off the surface into your eyes
or those of another person.

• Do not allow the laser light beam to impinge upon any

gas which can explode.

Spot size increases with distance from the probe tip as shown

(Spot Diameter measured at 90 % Energy)

OPERATING INSTRUCTIONS

Auto Power-down function

If unused for about 

15 seconds

, the meter will power-

down automatically. Press MEAS button switches to resume
power-on mode.

Turning the Power on

When the power is down, press MEAS 

less than 1 sec

button to turn on the meter.

The values and settings on the LCD return to what they

were before the power was last turned off.

When the power is down, press MEAS more than 1 sec,

the last record of MAX/MIN was erased.

 Display Back-Light Button

Press " " button to turn on or off the LCD Back-Light

during measurement, or while the display shows HOLD,
MAX or MIN.

Laser Sighting

Press " " button to on the " " annunciator. If " "

annunciator is on, press MEAS button and the laser beam
will turn on and " " annunciator will blink. Releasing
MEAS button turns off the laser beam.

ELECTRICAL

Temperature Range:

 

-4°F

 to 1022°F / 

-20°C

 to 550°C

Display Resolution:

 1°F , 0.5°C/1°C(Auto)

Accuracy:

 ±2% of reading or ±6°F/3°C, whichever is

greater @ 64.4 to 82.4°F (@ 18 to 28°C) ambient
operating temperature.

Temperature Coefficient:

 ±0.2% of reading or ±0.36°F/

0.2°C, whichever is greater, change in accuracy per °F/
°C change in ambient operating temperature above 82.4°F/
28°C or below 64.4°F/18°C.

Response Time: 

1 second

Spectral Response:

 6 to 14

µ

m nominal

Emissivity:

 0.10 to 1.00 by step of 0.01

Detection Element:

 Thermopile

Optical Lens:

 Fresnel Lens

Sighting:

 1-beam laser marker <1mW (class 2)

Field of View:

 100mmØ at 1000mm (3.9"Ø at 39.0")

OPERATING INSTRUCTIONS

 

  635

INFRARED THERMOMETER

WITH LASER SIGHTING

®

Selecting the Temperature Scale (°C or °F)

Readings are displayed in either degrees Celsius(°C) or

degrees Fahrenheit(°F). When the thermometer is turned
on, it is set to the temperature scale that was in use when the
thermometer was last turned off. To change the temperature
scale, in power down mode hold  the " " button (°F) [" "
button (°C)] down, and then press MEAS button until the °F
(°C) annunciator appear.

Continuous Measurement

1. In the power down mode, holding the MODE key down

and then pressing the MEAS key will put the meter into
the continuous measurement mode

, if HOLD was show,

please press MEAS button again to into the continuous
measurement.

2. When the MEAS key is pressed again to stop measuring

temperature and automatically hold the display reading,
the meter power down automatically after 

15 seconds.

NOTE:

 During the continuous measurement mode, the

HOLD indication does not appear.

NOTE:

 During the continuous measurement mode, the laser

sighting is not turned on.

MODE SELECTION AND OPERATION

Press MODE button switches the mode around the cycle 

ε

→ 

ALM Hi 

→ 

ALM Lo 

→ 

MAX 

→ 

MIN 

→ 

HOLD.

HOLD

:Releasing MEAS button stops measuring tem-

perature, the HOLD indication appears, and the
measured temperature is held.

ε

:

The thermal emissivity of the object set using
the 

▲ 

and 

▼ 

keys.

ALM Hi

:

The upper limit alarm temperature is set  using
the 

▲ 

and 

▼ 

keys. When the measured tem-

perature is exceeded the Hi setpoint, the beeper
emits a discontinuous pulse tone and "ALM Hi"
is displayed.

ALM Lo

:

The lower limit alarm temperature is set  using
the 

▲ 

and 

▼ 

keys. When the measured tem-

perature is blew the Lo setpoint, the beeper
emits a continuous pulse tone and "ALM Lo" is
displayed.

MAX

:The maximum temperature during measure-

ment is displayed. when measuring and also
pressing the mode key, this will enter MAX/
MIN/Present cycling.

MIN

:The minimum temperature during measure-

ment is displayed. when measuringand also
pressing the mode key, this will enter MAX/
MIN/Present cycling.