Page 1: ...Digital Thermometer Instruction manual Preparation in May 2018 The 2nd edition JAPAN BONKOTE CO LTD MCA 900Ⅱ ...

Page 2: ...sor unit SAT 1 A 2 Replacement sensor SC 006 A 3 How to measure iron tip temperature B Dip sensor SC 007 C Surface sensor SC 008 5 Explanation of each function 6 A Data hold function B MAX hold function C Selecting resolution function 6 Replacement of battery 7 7 Calibration 7 8 Guarantee and After sale service 7 A Guarantee B After sales service ...

Page 3: ...ine near heating equipment battery consumption will be faster than using it with normal temperature also it will have cause of damage and plastic melting of the body part Refrain from the place where would be direct sunshine for Display Considering condensation be careful to use this machine in environment that has rapid temperature change Working environment for this machine is 0 40 and under 80 ...

Page 4: ...ment range 0 1000 C Sensor Thermocouple K Measurement accuracy 0 3 rdg 2 C 23 C 5 C body only Display 3 5 Digit LCD Display Power source DC9V Battery 006P 1pcs Battery life 250hrs at continuous use of a brand new alkaline battery Working environment 0 40 C under 80 RH Dimension 80W x 48D x 156H mm including a soft case Weight 305 g including a battery C Display Temperature display degree Celsius H...

Page 5: ...andard is about 300 times measurement It is depends on each condition Replacement method is just pull it out and insert the new one Please make sure to put into correct pole Red color paint is pole A 3 How to measure iron tip temperature Put small amount of solder on the iron tip to measure the temperature Place the iron tip on the center of the sensor Refer to the left photo Make sure to place th...

Page 6: ...oom temperature will be indicated at the display of MCA 900Ⅱ Preparation is done Measurement is ready B Surface sensor SC 008 Measurement range Room temperature 500 C Connect the dipping sensor to MCA 900Ⅱ after confirming the polarity Turn on the power switch Approximate the room temperature will be indicated at the display of MCA 900Ⅱ Preparation is done Measurement is ready ...

Page 7: ...d function Push MAX key is turned on and highest measured value is indicated and hold Push MAX key again is turned off and release the function C Select resolution Push 0 1 1 key to select resolution at display Indication range of 0 1 resolution is 0 199 9 The value other than it OL will be indicated on a display For soldering work we recommend 1 resolution ...

Page 8: ...ification Calibration test reports and Traceability sheet come with the calibration If you have Calibration requests service please contact your dealer or us 8 Guarantee and after sales service A Guarantee Our products are shipped after sever factory tests inspections But if you find malfunction or defects due to problems in workmanship or transportation please contact with your dealer or us The g...