12
9 Vp-p
-4095
-4.5 V
4 Vp-p
-8191
-2 V
F4: Units
- Selects the amplitude units: peak-to-peak, RMS or dBm (sine waves only).
Note:
This option is shown when Ampl is selected.
F5: 50 OHM/HI-Z
- Selects the amplitude voltage value based on the two different impedance
termination (i.e. if connected to oscilloscope with 1MΩ input impedance,
generator will display the correct amplitude value for 1MΩ termination when HI-
Z is selected).
Note:
This option is shown when Ampl/Ofst is selected.
F5: Int Ref/Ext Ref
- Selects internal or external reference source (the external reference must be
connected to the rear panel Ref In connector).
Note
: This option is shown when Freq is selected.
WAVEFORM Key
Displays the waveforms available:
Figure 3.3 - Waveform Menu
F1: Sine
- Selects the sine waveform.
F2: Sqr
-
Selects the square waveform and displays the waveform duty cycle that can be
changed from 20% to 80% up to 10 MHz, 40% to 60% up to 30 MHz.
F3: Tri
-
Selects the triangle waveform and displays the waveform duty cycle that can be
changed from 0% to 100% up to 500 kHz, 10% to 90% up to 2 MHz, and 50% up to the
max frequency. The triangle maximum frequency is 5 MHz.
F4: Pulse
-
Selects the Pulse waveform and then displays the pulse menu. Refer to the PULSE
Menu section for details.
F5: ARB
-
Selects the arbitrary waveform and then displays the Arbitrary menu: