54
Fmin = 1/(100 S * Wavelength)
Rounding:
The value is rounded to 4 digits.
Command Type:
Setting or Query
Setting
Syntax:
[:SOURce]:FREQuency[:CW]<ws><frequency>[units]
[:SOURce]:FREQuency<ws>MINimum|MAXimum
Examples:
:FREQ 5KHZ
:FREQ 5E3
:FREQ MAXIMUM
:FREQ MIN
Query
Syntax:
[:SOURce]:FREQuency[:CW]?[<ws>MAXimum|MINimum]
Examples:
:FREQ?
:FREQ? MAX
Response:
NR3
Considerations:
1)
The MIN | MAX arguments should be used only in a Program Message that does NOT
contain Program Message Units specifying Arbitrary Point Rate or Wavelength, since the
MAXimum or MINimum value is calculated at the time the command is parsed.
2)
The MIN and MAX arguments refer to currently settable minimum or maximum.
3)
FIXed is alias for CW.
Amplitude
:SOURce:VOLTage[:AMPLitude] <p-p amplitude>
The amplitude command is used to set the peak-to-peak amplitude of the output waveform. Note that
the amplitude and the offset are limited by the relation
Peak Amp |Offset| ≤ 5V
Arguments
Type:
Numeric
Units:
V, mV, VPP, mVPP
Range:
10mV to 10V
Rounding:
1mV from 10mV to 999mV. 10mV from 1V to 10V.
Command Type:
Setting or Query
Setting
Syntax:
[:SOURce]:VOLTage:AMPLitude<ws><amplitude>[units]
[:SOURce]:VOLTage:AMPLitude<ws>MINimum|MAXimum
Examples:
:VOLT:AMPL 2.5
:VOLT:AMPL 2.5V
:VOLT:AMPL MAX
Query
Syntax:
[:SOURce]:VOLTage:AMPLitude?
www.calcert.com
1.800.544.2843
0
5
10
15
20
25
30