7
OPERATING MODES
Continuous
-
Output continuous at programmed parameters.
Triggered
-
Output quiescent until triggered by an internal or
external trigger, then one waveform cycle is generated to
programmed parameters, up to 2 MHz
Gate
-
Same as triggered mode, except waveform is executed for the
duration of the gate signal. The last cycle started is completed.
Trigger Source
-
Trigger source may be internal, external or manual.
Internal trigger rate 10us to 10s.
MODULATION CHARACTERISTICS
Amplitude Modulation
- Internal:
Sine signal of 1000 Hz
Variable modulation from 0% to 100% in 1% steps.
- External:
5 Vp-p for 100% modulation, 10 Kohms input impedance,
DC to 20 KHz bandwidth.
Frequency Modulation
- Internal:
Sine signal of 1000 Hz
- External:
5 Vp-p for 100% deviation, 10 Kohms input impedance,
DC to 20 KHz bandwidth.
SWEEP CHARACTERISTICS
Sweep Shape:
Linear and Logarithmic, up or down
Sweep Time:
10 ms to 50 s.
INPUTS AND OUTPUTS
Trigger In
-
TTL compatible.
-
Max. rate 2MHz.
-
Minimum width 50ns.
Sync Out
-
TTL pulse at programmed frequency, 50 ohms source impedance.
Modulation IN
-
5 Vp-p for 100% modulation .
-
10 K
input impedance. Dc to >20 KHz minimum bandwidth.
GENERAL
Dimensions
- 8.4 inch (213 mm) wide
- 3.5 inch (88 mm) high
- 8.3 inch (210 mm) deep
Weight
- Aprox 2.5 Kg.
Power
- 90 V-264V, 30 VA max
Temperature
- Operating - 0ºC to +50ºC, Non-operating - -10ºC to +70ºC
EMC
- According to EN55011 for radiated and conducted emissions.
Electrical Discharge Immunity
- According to EN55082
Safety Specifications
- According to EN61010
NOTE
Specifications are verified according to the performance check procedures in the technical manual.
Specifications not verified in the manual are either explanatory notes or general performance characteristics only.