13
The offset voltage has three ranges as follows:
Output Voltage Range
Constraints of Amp Offset
1.01 volt to 10.00 volts
(Vp-p)/2 + |offset| <= 5 volts
0.101 volt to 1 volt
(Vp-p)/2 + |offset| <= 0.5 volts
0.010 volt to 0.100 volt
(Vp-p)/2 + |offset| <= 0.05 volts
DUTY button
- When the Square or Triangle waveforms are selected, the variable SYMMETRY is
available. Change the waveform symmetry by using the cursor keys and rotary knob. If a
certain setting cannot be produced, the waveform generator will display a warning message.
Symmetry Menu
3.6.2 MODE Key
Selects the output mode: CONT (Continuous), TRIG (Triggered) and GATE (Gated).
To select the output mode, press MODE. Each additional MODE key operation selects TRIG, GATE or CONT
operating mode, Internal or External trigger source.
Mode Menu
Internal Trigger Menu