Code numbers are used to calibrate the
Omnitest
®
plus
meter for
accurate results. You must confirm the code number of the test strip
vial with the number shown on the meter, before using
Omnitest
®
plus
test strips.
Insert
Omnitest
®
plus
test strip to turn on the meter. The code
number will appear on the display for 3 seconds. Compare the code
number (example; C25) on the meter’s display with the code num-
ber on the test strip vial. If these two code numbers match, you can
start testing. If these two code numbers do not match, you must
code the meter according to the test strip vial code number.
15
14
Code the Meter
Without matching the code numbers, you will not get accu-
rate test results of your blood glucose level.
Caution
?
Press
button to increase the code number or
button to
decrease the code number. The number increases or decreases by
one digit every time you press
or
button while the code
number appears on the display. The code number increases or
decreases automatically and fast while holding to press
or
button.
If you missed to set the right code
number, insert the test strip again.
✔