2.3.1
Preparing holes for S
4
C screws with navigation
WARNING
Risk of injury to patients due to incorrect navigation!
►
Observe the combination specifications
►
To center the screw holes in the cortical bone layer for the self-tapping
S
4
C screws
Ø
3.5 mm with navigation, use the S
4
C navigated drill guide
Ø
3.5 mm
F
only together with S
4
C centering tool for S
4
C navigated drill
guide
P
.
To center the screw holes for the self-tapping S
4
C screws
Ø
4.0 mm with
navigation, use the S
4
C navigated drill guide
Ø
4.0 mm
G
only together
with S
4
C centering tool for S
4
C navigated drill guide
Q
.
S
4
C navigated drill guide
G
and centering tool
Q
have a purple marking on
the shaft for the self-tapping S4C screws
Ø
4.0 mm, see Fig. 1.
Fig. 1
Purple markings on S
4
C navigated drill guide and centering tool
To center the screw holes in the cortical layer without navigation, see
instructions for use of S
4
Cervical instruments (TA011984).
WARNING
Risk of damage to the spinal cord, nerve roots, adjacent intervertebral
space or soft tissue when inserting the centering tool without drill
guide!
►
Use the centering tool only with drill guide FW664R/FW665R.
CAUTION
Restriction of the function of the centering tool or inaccurate reading
of the depth gage!
►
Do not repoint the centering tool.
►
Replace blunt centering tools.
►
For mounting the reflective Brainlab marker spheres
4
onto Aesculap
star unit
H
, see Brainlab user manual.
►
Retract and hold locking sleeve
1
of the Aesculap star unit
H
against
the spring pressure in the direction of the arrow.
►
Push the Aesculap star unit
H
onto adapter
2
of the S
4
C drill guide
F
/
G
.
When doing so, ensure that the pin of the adapter
3
engages in the
recess on the star unit.
►
Release locking sleeve
1
.
Fig. 2
Mounting the Aesculap star unit on S
4
C drill guide
FW664R/FW665R
►
Prior to each use, check and/or validate the instrument with the
Brainlab instrument calibration matrix version 4, see Brainlab user
manual. When doing so, ensure that the centering tool for navigated
drill guide
P
/
Q
is removed.
►
To ensure that the camera has an unrestricted view of the reflective
marker spheres, loosen knob
6
on the S
4
C handle of the S
4
C drill guide
F
/
G
and rotate Aesculap star unit
H
to the desired position, see Fig. 2.
►
Once the desired position is reached, tighten knob
6
again.
►
Adjust the center punch depth by turning the depth stop 5 on the S
4
C
drill guide
F
/
G
. The maximum depth is 6 mm.
►
Insert S
4
C centering tool
P
/
Q
into S
4
C drill guide
F
/
G
.
►
Check the pre-set center punch depth with a caliper (e.g. AA845R).
►
To open the cortical bone, press S
4
C centering tool to the preset depth
under control with the Brainlab navigation system.
2.3.2
Drilling holes for the S
4
C screws
DANGER
Risk of injury from an incorrectly placed hole or a hole that is too
deep!
►
Do not sharpen the drill, as this would cause imprecise or incorrect
readings on the depth gauge.
►
Replace blunt drills with new ones.
The drill is applied with a S
4
C drill guide and drilled in either manually with
the drill handle (FJ839R) or with a motor system with the Aesculap Intra-
handpiece (e.g. GD450R/GD456R).
Assembling the drill and drill handle (for manual drilling only)
DANGER
Risk of damage to the spinal cord, nerve roots, adjacent intervertebral
space or soft tissue through incorrect drilling!
►
Use only the correct S
4
C drill guides to drill holes. Insert drill only
with the correct drill guide.
►
Before drilling, the pre-set drill length must be checked with a cal-
iper (e.g. AA845R, Caspar instrument for anterior cervical fusion).
Q
G
1
4
F/G
H
5
6
P/Q
2
3