5
Only process chemicals that have been tested and
approved (e.g. VAH or FDA approval or CE mark) and
which are compatible with the product’s materials
according to the chemical manufacturers’ recommen-
dations may be used for processing the product. All the
chemical manufacturer's application specifications
must be strictly observed. Failure to do so can result in
the following problems:
■
Optical changes of materials, e.g. fading or discol-
oration of titanium or aluminum. For aluminum,
the application/process solution only needs to be
of pH >8 to cause visible surface changes.
■
Material damage such as corrosion, cracks, fractur-
ing, premature aging or swelling.
►
Do not use metal cleaning brushes or other abra-
sives that would damage the product surfaces and
could cause corrosion.
►
Further detailed advice on hygienically safe and
material-/value-preserving reprocessing can be
found at www.a-k-i.org, link to Publications, Red
Brochure – Proper maintenance of instruments.
Disassembling the product before carry-
ing out the reprocessing procedure
►
Disassemble the product immediately after use, as
described in the respective instructions for use.
Prepare the double and single chisels
►
Slide cage
d
back and rotate counterclockwise by
90°.
►
Position guideway
e
on cage
d
in the chisel bridge,
see Fig. 3.
Fig. 3
Prepare the chisel guide
►
Push irrigation tube A
∅
7/I
∅
4 at least 20 mm
onto the end of the guide. In doing so, ensure that
guideway
f
is covered by the tube.
►
Adjust the safety stop to the central position
g
, see
Fig. 4.
Fig. 4
d
e
f
g