Azure Biosystems Azure 200 User Manual Download Page 14

Azure Imaging System User Manual

Page 13

 

9.   Sensitivity – see Section 3.2 for details. 

 

10.  Exposure – see Section 3.2 for details.

 11. 

UV Override – select this to customize UV light settings.

 

 

 •   Band Excision – requires overriding the safety interlock. Open the system door, then pull the 

override safety interlock switch.

    •  

Light On – 5 minutes (check box to automatically acquire image when time is expired). This step is 

necessary to warm up the UV table for the UV dye.

   

WARNING:

 ALWAYS WEAR APPROPRIATE PROTECTION WHEN EXPOSED TO UV LIGHT. The use 

of protective eyeglasses, mask, and/or gloves is strongly recommended.

 

 

12.   If you are interested in repositioning your sample before you take the image, use Live Mode

See Section 3.2 for details.

 

13.   Select          to capture the image.

 14. 

 The images will be displayed in the Gallery. The Gallery can be reached by clicking the tile icon. See 

Section 4.1 for instructions on using the Gallery.

 

15.  Select          to go back one screen.

3.6  Imaging Workflow for Protein Gel

 

 Use this for protein gel applications (e.g. polyacrylamide gel electrophoresis – PAGE).

 

1.    Remove the black blot tray and place sample in center of tray. Make sure to place the blot in the 

center of the tray.

 

2.   Close the Azure Imaging System door.

 

3.   Select Nucleic Acid Gel from the home screen.

 

4.   Select dyes in between one and four channels: red, green, blue, and grayscale. 

 

5.    Pick the name of the dye you are imaging. Note, if your dye does not appear on the list, contact Azure 

Technical Support. 

 

6.     Select  Auto Scan to produce an image with default settings. This is best to use when you don’t know 

what you expect to see on the blot. See Section 3.2 for additional details. 

 

7.   Select Manually Image to customize the settings.

 

8.   Sensitivity – See Section 3.2 for details. 

 

9.   Exposure – See Section 3.2 for details.

 

10.   If you are interested in repositioning your sample before you take the image, use Live Mode

See Section 3.2 for details.

 

11.   Select          to capture the image.

 

12.   The images will be displayed in the Gallery. The Gallery can be reached by clicking the tile icon. See 

Section 4.1 for instructions on using the Gallery.

 

13.  Select          to go back one screen.

Summary of Contents for Azure 200

Page 1: ...Azure Imaging Systems User Manual for the Azure 200 Azure 280 Azure 300 Azure 400 Azure 500 Azure 600 NIR RGB CHEMI WHITE LIGHT BLUE LIGHT UV Rev 20190919 ...

Page 2: ...any UV protection devices such as the safety switch on the light cabinet apparatus are working properly If not discontinue use until the device s are properly repaired Use only UV lamps in the transilluminator and ensure they are the correct size and voltage Electrical Safety Precautions Be sure to take proper precautions when handling any electrical equipment NEVER work on any live circuit fixtur...

Page 3: ...losed within the system enclosure and there are no viewing ports windows or openings to facilitate viewing of or exposure to radiation fields from direct impact reflection or leakage It is possible to override the safety switch which allows the user to turn on the light sources with the door open which may be useful for band excision in UV Maintenance Azure Biosystems 500 and 600 laser systems do ...

Page 4: ...orized agent Use of spare parts supplied by anyone other than Azure Biosystems Damage caused by accident or misuse Damage caused by disaster Corrosion caused by improper solvents or samples Voltage Setting Information The Azure Imaging System has a power supply that automatically chooses the correct voltage for your country or region CE Conformity The following Azure Imaging Systems models 200 280...

Page 5: ...B Input Devices 8 3 Capture Software Overview 9 3 1 The home page contains the following imaging options 9 3 2 Key definitions that apply to all imaging types 10 3 3 Imaging Workflow for Chemiluminescent Applications 11 3 4 Imaging Workflow for Fluorescent Applications 12 3 5 Imaging Workflow for Nucleic Acid Gel 12 3 6 Imaging Workflow for Protein Gel 13 3 7 Imaging Workflow for Custom Image 14 4...

Page 6: ...ter Wheel The 7 position filter wheel is automated It includes the standard Orange filter for most UV fluorescent applications and additional Red Green Blue and or IR filters for multichannel systems Integrated Computer The touch screen PC is completely integrated for system control and image acquisition Black Tray Stored in the front door the Black Tray also called Blot Tray is required for imagi...

Page 7: ... x 45 cm Weight 22 kg 49 lbs 1 2 Contacting Technical Support For questions about installation setup or general use of the system please contact support azurebiosystems com or call 925 307 7127 9am 4pm Pacific time Azure 600 Azure 500 Azure 400 Azure 300 Azure 280 Azure 200 Upgradable to 600 Upgradable to 600 Upgradable to 400 500 and 600 Upgradable to 300 400 500 and 600 NIR RGB CHEMI WHITE LIGHT...

Page 8: ...st light leaks 2 2 Connecting to Power The power entry module is located in the lower right on the back panel of the system Connect the power cord to a secure power outlet It is important to connect the system to a well grounded power source Azure Biosystems recommends that you employ a surge protecting power strip to protect against unexpected power surges that may cause damage to the electrical ...

Page 9: ...tware application If the software has been closed double click on the desktop icon Once the system is done initializing the software displays the gallery tab and is ready to use Please contact Azure technical support at support azurebiosystems com for questions about upgrading the software 2 5 Other USB Input Devices You may attach regulatory approved Windows OS supported USB keyboard USB mouse or...

Page 10: ... 3 Nucleic Acid Gel image nucleic acid blots See Section 3 5 for additional details 4 Protein Gels image protein blots See Section 3 6 for additional details 5 Custom select your preferred features from any of the other imaging modalities See Section 3 7 for additional details Other options include 6 Protocol selection use to save and load custom imaging protocols See Section 4 1 for details 7 Aut...

Page 11: ...e software to capture image for you using the full dynamic range of the camera There are four possible modes Underexpose software will optimize a short exposure time to the brightest signal on the blot Weaker signals may not be visualized in an underexposed image This is the fastest method Overexpose software will calculate an exposure time that saturates the brightest bands in the blot to potenti...

Page 12: ...f your dye does not appear on the list contact Azure Technical Support 7 Next select your imaging protocol Select Autoscan if you are not sure of your exposure time For single channel image chemi only The system defaults to 20 images with autoexpose Note this number is customizable in the settings tab It s like watching your blot develop incrementally before your eyes Select the best image in the ...

Page 13: ...ee Section 3 2 for details 10 If you are interested in repositioning your sample before you take the image use Live Mode See Section 3 2 for details 11 Select to capture the image 12 The images will be displayed in the Gallery The Gallery can be reached by clicking the tile icon See Section 4 1 for instructions on using the Gallery 13 Select to go back one screen 3 5 Imaging Workflow for Nucleic A...

Page 14: ...kflow for Protein Gel Use this for protein gel applications e g polyacrylamide gel electrophoresis PAGE 1 Remove the black blot tray and place sample in center of tray Make sure to place the blot in the center of the tray 2 Close the Azure Imaging System door 3 Select Nucleic Acid Gel from the home screen 4 Select dyes in between one and four channels red green blue and grayscale 5 Pick the name o...

Page 15: ...lect a focus value around the default setting Note For Azure 200 the focus does not need to be adjusted for the UV table or black tray 6 Select Exposure type See Section 3 2 for a detailed explanation of the exposure types 7 Add Dye add another set of options from numbers 3 6 above e g Excitation emission channel aperture focus and exposure times Can add up to four for the four channels red green ...

Page 16: ...en you return to the gallery 2 From the Print dropdown menu you can Print print to a local printer connected to the system or to a network printer Print Report print the image you re viewing as well as data about the image These include Exposure type Bin Level Gain Filter Calibration Software version System SN 3 Create a Multiplex image by merging between two and four channels 4 Create Signal Chan...

Page 17: ...these dyes choose any number of Chemi Fluorescent Protein Gel and Nucleic Acid Select File Saving Settings Can choose to enable autosave and or saving a publication image Simulation Mode allows you to run the software even if not connected to an Azure Imaging System Single Imaging Mode allows you to select the number of images in cumulative The default is 10 Auto Exposure Rapid capture will enable...

Page 18: ...s edit or create Custom Protocols Select Edit button to the right of existing protocols Or select light source name filter whether you want cumulative multiple or single images the default focus aperture and exposure and the live step and preview step settings ...

Page 19: ...ed The Azure Biosystems logo and Azure Biosystems are trademarks of the Company All other trademarks service marks and trade names appearing in this brochure are the property of their respective owners www azurebiosystems com info azurebiosystems com ...

Reviews: