B-5
INDEX Parameter name
Description
Sub-parameter
name
Access
attribute
Size
(Byte)
Range
Factory default
setting
48
FD_OFFSPEC_
MAP
Parameter that selects a condition
classified into this alarm category
(OFFSPEC). The same condition may
become active for more than one category
of the four alarm categories.
-
S-R/W
4
*1
0x00C00000
bit23/bit22
49
FD_MAINT_MAP
Parameter that selects a condition
classified into this alarm category
(MAINTENANCE). The same condition
may become active for more than one
category of the four alarm categories.
-
S-R/W
4
*1
0x0000E000
bit15/bit14/bit13
50
FD_CHECK_
MAP
Parameter that selects a condition
classified into this alarm category
(CHECK). The same condition may
become active for more than one category
of the four alarm categories.
-
S-R/W
4
*1
0x000000F
0bit7/bit6/bit5/
bit4
51
FD_FAIL_MASK
Parameter for the user to set multiple states
of this category (FAIL), which should not
be notified to the host through an alarm
parameter. Setting the bit to 1 inhibits
the notification, while setting the bit to 0
permits the notification.
-
S-R/W
4
*1
0
52
FD_OFFSPEC_
MASK
Parameter for the user to set multiple
states of this category (OFFSPEC), which
should not be notified to the host through
an alarm parameter. Setting the bit to 1
inhibits the notification, while setting the
bit to 0 permits the notification.
-
S-R/W
4
*1
0
53
FD_MAINT_
MASK
Parameter for the user to set multiple states
of this category (MAINTENANCE), which
should not be notified to the host through
an alarm parameter. Setting the bit to 1
inhibits the notification, while setting the
bit to 0 permits the notification.
-
S-R/W
4
*1
0
54
FD_CHECK_
MASK
Parameter for the user to set multiple
states of this category (CHECK), which
should not be notified to the host through
an alarm parameter. Setting the bit to 1
inhibits the notification, while setting the
bit to 0 permits the notification.
-
S-R/W
4
*1
0
55
FD_FAIL_ALM
Parameter for notifying the state change
of unmasked items of this alarm category
(FAIL) to the host system.
unacknowledged D-R/W
1
0: Undefined
1: Acknowledged
2: Unacknowledged
-
alarmstate
D-R
1
0: Undefined
1: Clear - reported
2: Clear - not reported
3: Active - reported
4: Active - not reported
-
timestamp
D-R
8
-
subcode
D-R
4
-
value
D-R
1
-
56
FD_OFFSPEC_
ALM
Parameter for notifying the state change
of unmasked items of this alarm category
(OFFSPEC) to the host system.
unacknowledged D-R/W
1
0: Undefined
1: Acknowledged
2: Unacknowledged
-
alarmstate
D-R
1
0: Undefined
1: Clear - reported
2: Clear - not reported
3: Active - reported
4: Active - not reported
-
timestamp
D-R
8
-
subcode
D-R
4
-
value
D-R
1
-
Summary of Contents for MagneW Plus+ MGG14C
Page 6: ...iv ...
Page 8: ...vi ...
Page 11: ...1 3 Figure1 3 System using a Foundation Fieldbus ...
Page 14: ...1 6 The following table is a summary of the functions of each of the keys ...
Page 26: ...2 6 ...
Page 42: ...4 6 ...
Page 78: ...A 10 ...
Page 105: ......
Page 108: ......
Page 110: ......