MEAS
MEM
LOG
SET
Welcome
9851/ 9651 Hygro-Thermometer Calibration Procedure
EXIT
LXW
EDIT
SAVE
View Cali Data
Long press
EXIT
LXW
MANUAL
RH Cali Mode Select
INPUT
Zcal : -7340019 E-4
Scal : 5399497 E-5
RH 32.8 %
EXIT
Cali
PRN
Temp:
REC
RH
WBT
DP
60.8%
17.4C
20.0C
Temp:25.5C
Input probe cali data
Press F2
Calibrate manually
Press F3
ABORT
LXW
ENTER
SAVE
Input Cali Data
RH Manual Calibration
RH 60.8%
RH Manual Calibration
RH 75.3 %
RH Manual Calibration
ABORT
Remarks:
1. Be sure to keep a calibration
note enclosed in the package
This's for future RH probe cali-
bration use
2. While keyin the calibration value,
3. Press ABORT anytime to exit without save or press END to
save and exit if you don't want to wait for 30 minutes.
4. The LCD value will flash while doing manual 33% and 75%
calibration until the procedure finished.
5. Suggest to calibrate at stable room temp. environment.
Flashing for 30 mins
Flashing for 30 mins
No need to use
salt bottles
Need to use
salt bottles
(Need to plug probe into salt bottles)
41
Note:
"Input cali.data" method is for RH
probe replacement.
"Manual calibration" method is for
RH probe replacement & old probe
re-calibration.
Zcal :
E-4
Scal : 5399497 E-5
-7340019
EXIT
75.3
32.8
(Put probe into corresponding salt bottles
for 30 mins before press F2 or F3)
END
ABORT
END
END
refer to the value indicated on
the note. The values are slope and offset value of each probes.
Temp. 23.5 C
Temp. 23.5 C
Temp. 23.5 C