CALIBRATION
PTS: 100.0%
TEMP: 27.0C
EXIT
LXW
ENTER
SAVE
RST
ATC
( Quit with save? )
Quit without save?
YES
NO
CALIBRATION
CALIBRATION
PTS:
75.0%
TEMP: 27.0C
EXIT
LXW
ENTER
SAVE
RST
ATC
Quit with save?
YES
NO
CALIBRATION
SAVE ENTER
CALIBRATION
Failure !!!
EXIT
LXW
RST
Ex. PTS's value is 75%
Remarks:
9861/9661 pH/mV Probe Calibration procedure-2
1.When you see "PTS:xx.x% , TEMP: xx.xC" shows on the
screen , that means the calibration of the point is done .
Press Enter to next calibration point or SAVE to stop cal
and store the calibration value .
2.You will see "Failure !!!" after you press YES to try to
save an unacceptable PTS range after calibration.
Under "Failure!!!" , press EXIT to exit calibration mode ,or
press RST to normal mode ( seeremark 3 for the details) .
Inactivate when pressing SAVE or ENTER after "Failure!!!" is
appeared.
3. RST stands for Reset , while proceeding any step, you
may stop calibration by pressing RST (F2) to back to
default setting.
Continuing from last page
Back to
Auto judging
next point cal.
Press EXIT ,"Quit without save?"
is appeared; Press SAVE,"Quit
with save ?" is appeared.
26