Installation
•
7
Chapter 2
d. Press the Configure button. In the
Configure Digidata 132x
dialog, press
the Detect button. The word "Detected" will appear and the OK button will
be enabled.
e. Press the Calibrate button. A dialog will pop up warning you to disconnect
any BNC cables from the Analog Outputs. Press OK, and in a couple of
seconds it should
say “Calibrated”
underneath the
Calibrate button.
Press the OK
buttons until all the
dialogs are closed.
We recommend
that the Digidata
1322A be
calibrated
whenever there is a
change in the
environment, such
as moving it to
another room or
installing it on
another computer.
f. The Digidata 1322A is now ready to perform experiments.