12
•
USE OF THE PATCH CLAMP – A TUTORIAL
Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.
Single Channel Recording (Model Cell)
Set the front panel controls of the Axopatch 200B as follows:
PIPETTE OFFSET:
About 5.0
PIPETTE CAPACITANCE COMP.
Minimum
(fully counterclockwise)
ZAP:
0.5 ms
SERIES RESISTANCE COMP. % PREDICTION:
0 %, OFF
SERIES RESISTANCE COMP. % CORRECTION:
0 %, OFF
SERIES RESISTANCE COMP. LAG:
1 µs
WHOLE CELL CAPACITANCE:
0 pF, OFF
SERIES RESISTANCE:
0 M
Ω
HOLDING COMMAND:
0 mV, x1, OFF
SEAL TEST:
ON
METER:
Set switch to V
TRACK
MODE:
TRACK
CONFIG.:
PATCH
OUTPUT GAIN:
α
= 10
LOWPASS BESSEL FILTER:
5 kHz
LEAK SUBTRACTION:
∞
M
Ω
, OFF
Pipette Offset Adjustment
Install the PATCH-1U model cell in BATH position into your headstage connector as
described in the
FUNCTIONAL CHECKOUT
section. Again, surround it with aluminum
foil grounded to the headstage ground connector. Connect the SCALED OUTPUT BNC to
one channel of your oscilloscope. Set the oscilloscope gain to 1 V/div and set its sweep
speed to 2 ms/div. Select line synchronization.
Using the TRACK Mode
You will see on the oscilloscope a rectangular, somewhat drooping pulse lasting about 8 ms
(10 ms for 50 Hz line frequency). This is the current driven through the 10 M
Ω
"electrode"
Summary of Contents for Axopatch 200B
Page 2: ......
Page 8: ...INFORMACION IMPORTANTE Axopatch 200B Copyright March 1997 1999 Axon Instruments Inc ...
Page 26: ......
Page 56: ......
Page 86: ......
Page 118: ......
Page 134: ......
Page 136: ......
Page 146: ......
Page 148: ......
Page 150: ......
Page 152: ......
Page 153: ...Circuit Diagrams Request Form ...