![Axon Axopatch 200B Theory Of Operation Download Page 121](http://html.mh-extra.com/html/axon/axopatch-200b/axopatch-200b_theory-of-operation_3040439121.webp)
SPECIFICATIONS
•
105
Chapter 9
Maximum Instrument Noise
Measured with minimal external noise sources (
i.e.
, radiated line frequency noise,
mechanical vibration), 8-pole Bessel filter.
PATCH
WHOLE CELL
β
= 1
β
= 0.1
Line freq.
Without holder:
& harmonics
0.005 pAp-p
0.005 pAp-p
0.005
pAp-p
0.1-100 Hz
0.030 pAp-p
0.50 pAp-p
1.60
pAp-p
0.1-1
kHz
0.015 pA rms
0.25 pA rms
0.75
pA rms
0.1-5
kHz
0.060 pA rms
0.65 pA rms
1.65
pA rms
0.1-10 kHz
0.130 pA rms
1.10 pA rms
3.00
pA rms
With holder:
0.1-10 kHz
0.145 pA rms
1.10 pA rms
3.00
pA rms
2
2
10
-29
-30
10
10 Hz
100 Hz
FREQUENCY
1 kHz
700 G
Ω
CURRENT NOISE DENSITY
(A / Hz)
10
-31
-32
10
Figure 23A.
Typical low frequency current noise spectrum, patch mode.
Summary of Contents for Axopatch 200B
Page 2: ......
Page 8: ...INFORMACION IMPORTANTE Axopatch 200B Copyright March 1997 1999 Axon Instruments Inc ...
Page 26: ......
Page 56: ......
Page 86: ......
Page 118: ......
Page 134: ......
Page 136: ......
Page 146: ......
Page 148: ......
Page 150: ......
Page 152: ......
Page 153: ...Circuit Diagrams Request Form ...