Axium Clinical Programmer User Manual Download Page 9

13

14

Leads Information Tab [Profile>Leads]

Lead 1

 through 

Lead 4

 are the default labels used to 

identify the implanted leads in the “Stim” Workspace. 

It is recommended that these names be changed into 

something more meaningful, for example the body 

region it covers.

•   

Target Name:

 for each of the implanted leads, 

enter the body region covered (text field).

For each of the leads enter the Lot and Model 

number:
• 

  Lot #: 

Enter the Lot number found on the lead 

packaging.

•  

Model #:

 Enter the lead Model number.

System Information Tab [Profile>System]

From the system tab the following system parameters 

can be managed: 

•  

 Periodic Impedance Interval: 

Set the 

periodicity with which you want the system to 

measure lead impedance.

• 

  Follow-up Period: 

A calculated field which 

displays the recommended follow-up time 

based on the programmed settings. It is an 

indicator of when the programmer will run out 

of memory and will begin to overwrite old data 

for impedance measurements.

•   

Ramp Duration:

 Ramp duration is how long 

it takes for the NS to reach the requested 

amplitude. If set to 8 seconds, the NS will 

take 8 seconds to get from 0 to the requested 

amplitude when a lead is switched from not 

enabled to enabled. Ramping also occurs when 

the step between the current amplitude and the 

next amplitude is greater than 100 mV.

•   

Magnet Magnet Stim On/Off Time: 

Allows 

you to control how long it takes before a 

magnet held over the device switches off or 

back on delivered therapy.

Stim Workspace

Press the “Stim” tab to access the Stimulation Settings Workspace. The Stimulation Settings Workspace 

is divided into five tabs which are used to:

•  Activate (turn on) up to four leads
•  Adjust electrode configurations
•  Measure impedance
•  Set nominal values to begin stimulation
•  Perform trial mapping
•  Confirm the response and sensation of specific body regions to be stimulated

Stim Tabs [Stim>Target Name]

The Stim tabs are the main tabs from which therapy is 

controlled and programmed. This can be done either 

temporarily (testing) or permanently. 

•   

Select Group:

 Select the group for which you 

want to change the stimulation settings.

NOTE:

 in the Group Workspace, up to four 

different groups can be defined each with 

their own stimulation parameters. A group can 

be linked for example to a specific activity or 

posture. Refer to the Group Workspace section 

in this manual for more information.

•   

Select the Tab:

 Labeled with the target 

for which you want to adapt stimulation 

parameters. (In the sample screen e.g. “L 

Foot”)

NOTE:

 There are up to four tabs that can be 

labeled with the body region in which stimulation 

with the corresponding lead is targeting (defined 

in Profile>Leads). For each body region (lead) 

stimulation can be adjusted independently.

•  

Patient Refresh Impedance:

 This allows the impedance to be remeasured and used  

   for therapy on the Patient Programmer whenever the patient turns a Lead Enable from   

   OFF to ON.

 NOTE: 

the programmed Pulse Amplitude will be adjusted based on the  

   updated impedance to maintain a constant therapy voltage.

Summary of Contents for Clinical Programmer

Page 1: ...thorized duplication or use is prohibited AXIUM SPINALMODULATION and the Man in Circle Design are trademarks and or registered trademarks of Spinal Modulation Inc Clinical Programmer Model MN20700 03 User Manual Authorization to affix the CE mark granted in 2011 ...

Page 2: ... Workspace Navigation 9 Temporary and Permanent Programming 10 More on Editing Text Fields 10 USING THE WORKSPACES 10 Profile Workspace 10 Patient Information Tab Profile Patient 11 Clinic Information Tab Profile Clinic 12 Stimulator Information Tab Profile NS 12 Leads Information Tab Profile Leads 13 System Information Tab Profile System 13 Stim Workspace 14 Stim Tabs Stim Target Name 14 Impedanc...

Page 3: ...ds and stimulator hardware used differ the programmer hardware and instructions for programming the TNS and INS devices are the same NOTE In this manual the general abbreviation NS is used for information which applies to both TNS and INS In all other cases the specific abbreviations TNS or INS are used For specific description of the TNS and INS system components and implant procedures refer to t...

Page 4: ... concern regarding the proper function of the Spinal Modulation NS System please contact your Spinal Modulation representative RF OPERATING FREQUENCIES Nearby equipment emitting strong magnetic fields can interfere with RF communication even if the other equipment complies with CISPR emission requirements The operating characteristics are as follows MICS band 402 405 MHz The effective radiated pow...

Page 5: ...cycles with normal use Connect the Clinical Programmer to the Charger and attach to an outlet regularly to keep it charged Programmer Power Up Turn the Clinical Programmer ON by pressing the button The Main Menu will be displayed NOTE If the Clinical Programmer screen does not turn on follow the instructions for charging the battery and try again Main Menu The Main Menu displays three primary func...

Page 6: ... telemetry connection is cancelled 6 Move the Clinical Programmer magnet over the NS device in a circular motion to connect The indicator status bar on the bottom left of the screen will display Connected if the connection attempt is successful If the Programmer could not communicate with the NS device an error message will appear and Disconnected will be displayed in the status bar NOTE If after ...

Page 7: ...e programmed NS settings NOTE When programming is complete select the Exit button to conserve power Temporary and Permanent Programming Whenever a change is made to a parameter value or other data field while the NS is within telemetry range this value immediately becomes temporarily active The corresponding value or data selection appears in a red bold underlined font NOTE The temporarily active ...

Page 8: ...ng the áü button near the space bar allows the use of accented characters Clinic Information Tab Profile Clinic Enter or modify the physician and clinic information in the text fields provided Physician Name Clinic Name Clinic After Hours Contact Phone Number Clinic Phone Number Clinic Email Clinic Address Stimulator Information Tab Profile NS The NS tab provides a summary of information related t...

Page 9: ...the device switches off or back on delivered therapy Stim Workspace Press the Stim tab to access the Stimulation Settings Workspace The Stimulation Settings Workspace is divided into five tabs which are used to Activate turn on up to four leads Adjust electrode configurations Measure impedance Set nominal values to begin stimulation Perform trial mapping Confirm the response and sensation of speci...

Page 10: ...st be valid prior to activating the lead A valid lead configuration must include at least one positive and one negative electrode Lead Enable must be ON before the amplitude can be increased from 0 µA Location Enter the spinal level where stimulation therapy is delivered by this lead Pulse Parameters To select and change pulse parameters first press the desired increment level Fine Medium Coarse A...

Page 11: ...to the programmed durations when out of session or in session in the Group Workspace this allows the clinician to evaluate and optimize On Off durations When in session in the Stim Workspace Periodic Stimulation will be inactive stimulation will remain ON to allow the clinician to optimize pulse parameters Rework Tabs Utility Rework This screen provides one button automation of the normal TNS rewo...

Page 12: ...tient Programmer s Programmer Setup screen select the desired language then press the Bind Stimulator button Follow the sequence of prompts to enter the target NS serial number and complete the Binding process The Programmer will reboot upon completion Stimulator Unbinding On a bound Patient Programmer press the Unbind Stimulator button to remove the binding This will free the Patient Programmer t...

Page 13: ... of the test level Mains power quality should be that of a typical commercial or home environment Power frequency 50 60 Hz magnetic field IEC 61000 4 8 3 A m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial hospital or home environment Guidance and Manufacturer s Declarations GUIDANCE AND MANUFACTURER S DECLARATION Electromagnetic Emi...

Page 14: ...cy range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people Recommended separation distances between portable and mobile RF communications equipment and the Spinal Modulation Neurostimulation System The Spinal Modulation Neurostimulation System is intended for use in an electromagne...

Page 15: ... L C8 R C8 Magnet Turnoff Time Off 1 s 2 s 3 s 4 s 5 s 6 s 7 s 8 s 9 s 10 s 15 s 20 s 25 s 30 s 40 s 50 s 1 min 70 s 80 s 90 s 100 s 110 s 2 min Ramp Duration 1 s 2 s 3 s 4 s 5 s 6 s 7 s 8 s Appendix II Troubleshooting Pop up Messages Condition Buttons Resolution All stimulation has been turned OFF All stimulation turned off due to data corruption OK Contact your Spinal Modulation Representative C...

Page 16: ...FF Stimulation can be turned off either by the All Stim OFF software button on the INS programmer or the Off switch on the TNS OK Stimulator battery has reached End of Service EOS Stimulation has been turned OFF permanently Battery has reached EOS voltage Stimulation is disabled in order to preserve power for RF communication OK Schedule replacement of the Stimulator Pop up Messages Cont Condition...

Reviews: