Precautions

The following precautions should be taken to avoid damage to the Clinical Programmer and to ensure 

proper function:
    •   Do not drop or mishandle the Clinical Programmer. Physical damage to the Clinical Programmer may 

impair its function.

    •   Do not spill fluids on or wash the Clinical Programmer. Excessive moisture may impair its function. If 

cleaning is necessary, remove soil with a soft damp cloth.

    •   Do not use abrasive or caustic cleaning products on the Clinical Programmer.
    •   Do not attempt to open the case for the Clinical Programmer. Attempts to open the case may expose 

the Clinical Programmer to elements that alter its function.

    •   The Clinical Programmer has an internal magnet. Keep the Clinical Programmer away from any 

credit cards, hard drives, or magnetic storage devices as it may demagnetize them.

    •   Do not operate the Clinical Programmer outside the specified temperature range of 5°C to 40°C. 

 

Rapid temperature changes may affect proper device operation.

    •    Do not store the Clinical Programmer outside the specified temperature range of -10°C to 50°C.
    •   Do not leave the Clinical Programmer in a car or other places where temperatures can exceed 50°C.
    •   Do not burn or otherwise dispose of the Clinical Programmer. Fire may cause the internal battery to 

explode.

    •   Do not allow unauthorized use of the Clinical Programmer to avoid injury to patients.
    •   The NS device can only be programmed using Spinal Modulation’s Clinical or Patient Programmer. 

Do not try to use any other manufacturer’s device to program it.

    •   Do not use the Clinical Programmer or NS in the presence of explosive or flammable gases as this 

may cause serious injury.

    •   Do not use the Programmer Charger if the power cord is damaged, excessively worn or frayed. This 

may cause injury or damage the Programmer.

    •   Frequent programming of the implanted device will cause the battery to deplete faster. Avoid 

unnecessary programming.

    •   If there is any concern regarding the proper function of the Spinal Modulation NS System, please 

contact your Spinal Modulation representative.

RF OPERATING FREQUENCIES

Nearby equipment emitting strong magnetic fields can interfere with RF communication, even if the other 

equipment complies with CISPR emission requirements. The operating characteristics are as follows: 

MICS band: 402-405 MHz

The effective radiated power is below the limits as specified in
Europe: EN ETSI 301 839-2
USA FCC 47 CFR Part 95; 95.601-95.673 Subpart E, 95.1201-95.1219
FCC ID: Y8L-MN

2

0700

-03

 

This device may not interfere with stations operating in the 400.150–406.000 MHz band in the 

Meteorological Aids, Meteorological Satellite, and Earth Exploration Satellite Services and must accept 

any interference received, including interference that may cause undesired operation. 

Clinical Programmer System Overview

The Axium Clinical Programmer allows you to establish two-way communication with the patient’s NS 

device for querying and programming.

It is a portable, hand-held device that can be plugged into a power outlet or be powered by an internal 

battery. The battery is rechargeable using the power supply provided and a power outlet.

The Clinical Programmer System includes:
    •  Clinical Programmer with Stylus
    •  Programmer Charger
    •  External Auxiliary Magnet
    •  Programmer Carrying Case
    •  Clinical Programmer User Manual (this document)

Clinical Programmer Features

With the Clinical Programmer, you can:
    •  Turn OFF all stimulation.
    •  Turn stimulation ON for up to four leads and measure lead impedance.
    •  Change stimulation settings for each lead.
    •  Configure Patient Controlled Therapy settings for each lead.
    •   Enter patient and lead identification information, clinician and clinic name and contact information, 

and clinician’s notes.

    •   Create and name groups of stimulation sets with each group containing up to four leads with different 

settings on each lead. 

    •  Perform a real time trial (test) to assess the patient stimulation response for each lead.
    •  Acquire identification, diagnostic, and historic information about the NS device.

Clinical Programmer

Stylus

Programmer Charger

Summary of Contents for Clinical Programmer

Page 1: ...thorized duplication or use is prohibited AXIUM SPINALMODULATION and the Man in Circle Design are trademarks and or registered trademarks of Spinal Modulation Inc Clinical Programmer Model MN20700 03 User Manual Authorization to affix the CE mark granted in 2011 ...

Page 2: ... Workspace Navigation 9 Temporary and Permanent Programming 10 More on Editing Text Fields 10 USING THE WORKSPACES 10 Profile Workspace 10 Patient Information Tab Profile Patient 11 Clinic Information Tab Profile Clinic 12 Stimulator Information Tab Profile NS 12 Leads Information Tab Profile Leads 13 System Information Tab Profile System 13 Stim Workspace 14 Stim Tabs Stim Target Name 14 Impedanc...

Page 3: ...ds and stimulator hardware used differ the programmer hardware and instructions for programming the TNS and INS devices are the same NOTE In this manual the general abbreviation NS is used for information which applies to both TNS and INS In all other cases the specific abbreviations TNS or INS are used For specific description of the TNS and INS system components and implant procedures refer to t...

Page 4: ... concern regarding the proper function of the Spinal Modulation NS System please contact your Spinal Modulation representative RF OPERATING FREQUENCIES Nearby equipment emitting strong magnetic fields can interfere with RF communication even if the other equipment complies with CISPR emission requirements The operating characteristics are as follows MICS band 402 405 MHz The effective radiated pow...

Page 5: ...cycles with normal use Connect the Clinical Programmer to the Charger and attach to an outlet regularly to keep it charged Programmer Power Up Turn the Clinical Programmer ON by pressing the button The Main Menu will be displayed NOTE If the Clinical Programmer screen does not turn on follow the instructions for charging the battery and try again Main Menu The Main Menu displays three primary func...

Page 6: ... telemetry connection is cancelled 6 Move the Clinical Programmer magnet over the NS device in a circular motion to connect The indicator status bar on the bottom left of the screen will display Connected if the connection attempt is successful If the Programmer could not communicate with the NS device an error message will appear and Disconnected will be displayed in the status bar NOTE If after ...

Page 7: ...e programmed NS settings NOTE When programming is complete select the Exit button to conserve power Temporary and Permanent Programming Whenever a change is made to a parameter value or other data field while the NS is within telemetry range this value immediately becomes temporarily active The corresponding value or data selection appears in a red bold underlined font NOTE The temporarily active ...

Page 8: ...ng the áü button near the space bar allows the use of accented characters Clinic Information Tab Profile Clinic Enter or modify the physician and clinic information in the text fields provided Physician Name Clinic Name Clinic After Hours Contact Phone Number Clinic Phone Number Clinic Email Clinic Address Stimulator Information Tab Profile NS The NS tab provides a summary of information related t...

Page 9: ...the device switches off or back on delivered therapy Stim Workspace Press the Stim tab to access the Stimulation Settings Workspace The Stimulation Settings Workspace is divided into five tabs which are used to Activate turn on up to four leads Adjust electrode configurations Measure impedance Set nominal values to begin stimulation Perform trial mapping Confirm the response and sensation of speci...

Page 10: ...st be valid prior to activating the lead A valid lead configuration must include at least one positive and one negative electrode Lead Enable must be ON before the amplitude can be increased from 0 µA Location Enter the spinal level where stimulation therapy is delivered by this lead Pulse Parameters To select and change pulse parameters first press the desired increment level Fine Medium Coarse A...

Page 11: ...to the programmed durations when out of session or in session in the Group Workspace this allows the clinician to evaluate and optimize On Off durations When in session in the Stim Workspace Periodic Stimulation will be inactive stimulation will remain ON to allow the clinician to optimize pulse parameters Rework Tabs Utility Rework This screen provides one button automation of the normal TNS rewo...

Page 12: ...tient Programmer s Programmer Setup screen select the desired language then press the Bind Stimulator button Follow the sequence of prompts to enter the target NS serial number and complete the Binding process The Programmer will reboot upon completion Stimulator Unbinding On a bound Patient Programmer press the Unbind Stimulator button to remove the binding This will free the Patient Programmer t...

Page 13: ... of the test level Mains power quality should be that of a typical commercial or home environment Power frequency 50 60 Hz magnetic field IEC 61000 4 8 3 A m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial hospital or home environment Guidance and Manufacturer s Declarations GUIDANCE AND MANUFACTURER S DECLARATION Electromagnetic Emi...

Page 14: ...cy range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people Recommended separation distances between portable and mobile RF communications equipment and the Spinal Modulation Neurostimulation System The Spinal Modulation Neurostimulation System is intended for use in an electromagne...

Page 15: ... L C8 R C8 Magnet Turnoff Time Off 1 s 2 s 3 s 4 s 5 s 6 s 7 s 8 s 9 s 10 s 15 s 20 s 25 s 30 s 40 s 50 s 1 min 70 s 80 s 90 s 100 s 110 s 2 min Ramp Duration 1 s 2 s 3 s 4 s 5 s 6 s 7 s 8 s Appendix II Troubleshooting Pop up Messages Condition Buttons Resolution All stimulation has been turned OFF All stimulation turned off due to data corruption OK Contact your Spinal Modulation Representative C...

Page 16: ...FF Stimulation can be turned off either by the All Stim OFF software button on the INS programmer or the Off switch on the TNS OK Stimulator battery has reached End of Service EOS Stimulation has been turned OFF permanently Battery has reached EOS voltage Stimulation is disabled in order to preserve power for RF communication OK Schedule replacement of the Stimulator Pop up Messages Cont Condition...

Reviews: