GUIDANCE AND MANUFACTURER’S DECLARATION

Electromagnetic Emissions

The Spinal Modulation Neurostimulator System is intended for use in the electromagnetic environment 

specified below. The customer or the user of the Spinal Modulation Neurostimulator System should 

assure that it is used in such an environment.

Immunity

IEC 60601 Test Level

Compliance 

Level

Electromagnetic 

Environment Guidance

Electrostatic 

discharge (ESD)

IEC 61000-4-2

± 6 kV contact

± 8 kV air

Floors should be wood, 

concrete or ceramic tile. 

If floors are covered with 

synthetic material, the relative 

humidity should be at least 

30%.

Electrical fast

transient/burst

IEC 61000-4-4

± 2 kV for power supply lines

± 1 kV for input/output lines

Pass

Mains power quality should be 

that of a typical commercial or 

home environment

Surge

IEC 61000-4-5

± 1 kV line(s) to line(s)

± 2 kV line(s) to earth

Mains power quality should be 

that of a typical commercial or 

home environment

Voltage dips, 

short interruptions 

and voltage 

variations on 

power supply

input lines

IEC 61000-4-11

<5% UT (>95% dip in UT)

for 0.5 cycle

40% UT (60% dip in UT)

for 5 cycles 

70% UT (30% dip in UT) 

for 25 cycles

<5% UT (>95% dip in UT) 

for 5 s

NOTE UT is the a.c. mains 

voltage prior to application of 

the test level.

Mains power quality should be 

that of a typical commercial or 

home environment

Power frequency 

(50/60 Hz) 

magnetic field

IEC 61000-4-8

3 A/m

Power frequency magnetic 

fields should be at levels 

characteristic of a typical 

location in a typical 

commercial, hospital, or home 

environment.

Guidance and Manufacturer’s Declarations 

  

 

 

GUIDANCE AND MANUFACTURER’S DECLARATION

Electromagnetic Emissions

The Spinal Modulation Neurostimulator System is intended for use in the electromagnetic environment 

specified below. The customer or the user of the Spinal Modulation Neurostimulator System should 

assure that it is used in such an environment.

Emissions test

Compliance Electromagnetic Environment – Guidance

RF Emissions 1

Group 2

The Spinal Modulation Neurostimulator System must emit 

electromagnetic energy in order to perform its intended 

function. Nearby electronic equipment may be affected.

RF emissions

CISPR 11

Class B

Harmonic emissions

IEC 61000-3-2

Class B

Voltage fluctuations/

flicker emissions

IEC 61000-3-3

Complies

The Spinal Modulation Neurostimulator System is suitable for 

use in all establishments, including domestic establishments 

and those directly connected to the public low voltage power 

supply network that supplies buildings used for domestic 

purposes.

CISPR 14-1

Complies

The Clinical Programmer is not intended to be connected to 

other equipment except the Programmer Charger.

Summary of Contents for Clinical Programmer

Page 1: ...thorized duplication or use is prohibited AXIUM SPINALMODULATION and the Man in Circle Design are trademarks and or registered trademarks of Spinal Modulation Inc Clinical Programmer Model MN20700 03 User Manual Authorization to affix the CE mark granted in 2011 ...

Page 2: ... Workspace Navigation 9 Temporary and Permanent Programming 10 More on Editing Text Fields 10 USING THE WORKSPACES 10 Profile Workspace 10 Patient Information Tab Profile Patient 11 Clinic Information Tab Profile Clinic 12 Stimulator Information Tab Profile NS 12 Leads Information Tab Profile Leads 13 System Information Tab Profile System 13 Stim Workspace 14 Stim Tabs Stim Target Name 14 Impedanc...

Page 3: ...ds and stimulator hardware used differ the programmer hardware and instructions for programming the TNS and INS devices are the same NOTE In this manual the general abbreviation NS is used for information which applies to both TNS and INS In all other cases the specific abbreviations TNS or INS are used For specific description of the TNS and INS system components and implant procedures refer to t...

Page 4: ... concern regarding the proper function of the Spinal Modulation NS System please contact your Spinal Modulation representative RF OPERATING FREQUENCIES Nearby equipment emitting strong magnetic fields can interfere with RF communication even if the other equipment complies with CISPR emission requirements The operating characteristics are as follows MICS band 402 405 MHz The effective radiated pow...

Page 5: ...cycles with normal use Connect the Clinical Programmer to the Charger and attach to an outlet regularly to keep it charged Programmer Power Up Turn the Clinical Programmer ON by pressing the button The Main Menu will be displayed NOTE If the Clinical Programmer screen does not turn on follow the instructions for charging the battery and try again Main Menu The Main Menu displays three primary func...

Page 6: ... telemetry connection is cancelled 6 Move the Clinical Programmer magnet over the NS device in a circular motion to connect The indicator status bar on the bottom left of the screen will display Connected if the connection attempt is successful If the Programmer could not communicate with the NS device an error message will appear and Disconnected will be displayed in the status bar NOTE If after ...

Page 7: ...e programmed NS settings NOTE When programming is complete select the Exit button to conserve power Temporary and Permanent Programming Whenever a change is made to a parameter value or other data field while the NS is within telemetry range this value immediately becomes temporarily active The corresponding value or data selection appears in a red bold underlined font NOTE The temporarily active ...

Page 8: ...ng the áü button near the space bar allows the use of accented characters Clinic Information Tab Profile Clinic Enter or modify the physician and clinic information in the text fields provided Physician Name Clinic Name Clinic After Hours Contact Phone Number Clinic Phone Number Clinic Email Clinic Address Stimulator Information Tab Profile NS The NS tab provides a summary of information related t...

Page 9: ...the device switches off or back on delivered therapy Stim Workspace Press the Stim tab to access the Stimulation Settings Workspace The Stimulation Settings Workspace is divided into five tabs which are used to Activate turn on up to four leads Adjust electrode configurations Measure impedance Set nominal values to begin stimulation Perform trial mapping Confirm the response and sensation of speci...

Page 10: ...st be valid prior to activating the lead A valid lead configuration must include at least one positive and one negative electrode Lead Enable must be ON before the amplitude can be increased from 0 µA Location Enter the spinal level where stimulation therapy is delivered by this lead Pulse Parameters To select and change pulse parameters first press the desired increment level Fine Medium Coarse A...

Page 11: ...to the programmed durations when out of session or in session in the Group Workspace this allows the clinician to evaluate and optimize On Off durations When in session in the Stim Workspace Periodic Stimulation will be inactive stimulation will remain ON to allow the clinician to optimize pulse parameters Rework Tabs Utility Rework This screen provides one button automation of the normal TNS rewo...

Page 12: ...tient Programmer s Programmer Setup screen select the desired language then press the Bind Stimulator button Follow the sequence of prompts to enter the target NS serial number and complete the Binding process The Programmer will reboot upon completion Stimulator Unbinding On a bound Patient Programmer press the Unbind Stimulator button to remove the binding This will free the Patient Programmer t...

Page 13: ... of the test level Mains power quality should be that of a typical commercial or home environment Power frequency 50 60 Hz magnetic field IEC 61000 4 8 3 A m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial hospital or home environment Guidance and Manufacturer s Declarations GUIDANCE AND MANUFACTURER S DECLARATION Electromagnetic Emi...

Page 14: ...cy range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people Recommended separation distances between portable and mobile RF communications equipment and the Spinal Modulation Neurostimulation System The Spinal Modulation Neurostimulation System is intended for use in an electromagne...

Page 15: ... L C8 R C8 Magnet Turnoff Time Off 1 s 2 s 3 s 4 s 5 s 6 s 7 s 8 s 9 s 10 s 15 s 20 s 25 s 30 s 40 s 50 s 1 min 70 s 80 s 90 s 100 s 110 s 2 min Ramp Duration 1 s 2 s 3 s 4 s 5 s 6 s 7 s 8 s Appendix II Troubleshooting Pop up Messages Condition Buttons Resolution All stimulation has been turned OFF All stimulation turned off due to data corruption OK Contact your Spinal Modulation Representative C...

Page 16: ...FF Stimulation can be turned off either by the All Stim OFF software button on the INS programmer or the Off switch on the TNS OK Stimulator battery has reached End of Service EOS Stimulation has been turned OFF permanently Battery has reached EOS voltage Stimulation is disabled in order to preserve power for RF communication OK Schedule replacement of the Stimulator Pop up Messages Cont Condition...

Reviews: