15.7. Function for calibration with external weight (CALIb)
Calibration with external weight should be performed if scale accuracy is not
satisfactory. Calibration weight stated in technical data table for the scale (or of
better accuracy) should be used.
Operations sequence:
Press
MENU
key to display user
functions, shown one by one in
loop.
Press
→
T
←
key when
CALIbr
function appears.
The following options will be
displayed:
-
CAL oFF
–
switch
internal
calibration off
-
CAL on
–
calibration
with
external weight
out
– leave without changes
Press
→
T
←
key when
CAL on
option appears.
Wait for writing zero to the scale.
When
LOAD
message appears put
calibration weight on the pan.
Wait until internal calibration is
finished and zero indication is
displayed.
- CAL -
0.00g
LOAD
m
MENU
T
T
0.00g
0.00g
CALIbr
CAL on
CAL oFF
...
m