15.3. Pieces counting function (PCS)
The function allows counting
identical pieces, e.g. pills or buttons
included into weighed portion.
Measurement is performed in two
phases:
- first phase – calculating the mass of
single item basing on a sample
containing
defined
amount
of
pieces:
5, 10, 20, 50, 100, 200
or
500
pieces,
- second phase – counting pieces in
weighed portion.
The function has the following
options:
-
PCS OFF
– switch the function off
-
PCS on
– switch the function on
-
PCS ...
– recover last used unitary
mass,
-
PCS 5, 10, 20
, ... , 500 – amount of
pieces in a sample,
-
PCS SEt
– set any amount of pieces
in a sample,
-
PCS MJ
– set unitary mass directly,
-
out
– leave menu without changes.
Comments:
1.
It is recommended that mass of
single piece is greater than reading
unit and mass of sample used in
first phase is bigger than 100
reading units.
2. Err-PCS message indicates that a
sample was not put on a pan or a
mass of single piece is less than one
reading unit (it is possible to count
pieces knowing that errors will be
bigger).
0 pcs
0.00 g
12.05 g
23.04 g
PCS
PCS oFF
PCS on
5
5
12
12 pcs
MENU
T
T
T