Operation sequence:
Application options:
-
Activation
–
move to weighing with summing,
-
Zero threshold
–
zero signalling threshold,
-
MIN threshold
–
Minimum threshold signalling,
-
MAX threshold
–
Maximum threshold signalling,
-
Mode
–
working mode:
<signaller>
indication mode (chart on previous
site)
,
<impulse>
impulses and sound signal (batching
mode chart on previous site),
-
Printout
–
threshold printout,
-
Schortcut
-
fast access key
selection: F1, F2,…
or F5.
Actions order:
No load (load smaller than zero threshold)
-
no signalization
Put weight.
Case 1:
Load under
MIN
threshold.
- balance signalizes to small value
–
MIN
.
Case 2:
Load above
MIN
threshold but under
MAX
- balance signalizes good value
–
OK (in
Impulse
mode additional short sound signal occurs)
Case 3:
Load above
MAX
.
- balance signalizes too big value
–
MAX
(in
Impulse
mode additional long sound signal occurs).
APPLICATIONS
...
Threshold
...
Exit
THRESHOLD
1. Activation
2. Zero threshold
0.0010 g
3. MIN threshold
1.0000 g
4. MAX threshold
10.0000 g
5. Mode:
<signaller><impulse>
6. Printout
7. Shortcut:
F5
8. Exit
Max ... Min ... e= ... d= .. I
AUT
0.0000
g
0%
100%
Max ... Min ... e= ... d= .. I
MIN
AUT
0.1002
g
0%
100%
Max ... Min ... e= ... d= .. I
OK
AUT
5.0002
g
0%
100%
Max ... Min ... e= ... d= .. I
MAX
AUT
20.0002
g
0%
100%