Blade V2 Manual
Model
Acceptable
Tolerance (±)
Calibration
Mass
Recommended
Accuracy Class
BL2-50
0.02g
50g
OIML M2
BL2-100
0.02g
100g
OIML M2
BL2-400
0.2g
200g
OIML M2
VIII. Adjustment
1. Press the key to turn the scale on, then immediately
press and hold the M key until “CAL” is displayed
then release.
2.
Press the M key again. The display will flash “CAL”
followed by the required calibration weight (see
above).
3. Place the required calibration weight on the center
of the weighing platform.
4.
Once the weight is confirmed, the display will show
“PASS” then return to normal weighing mode.
NOTE:
Adjustment using the incorrect mass will
result in inaccurate readings.
IX. Troubleshooting
•
Problem:
Scale will not turn on
•
Solution:
Check the batteries and battery contacts.
•
Problem:
Displayed weight fluctuates randomly
•
Solution:
Make sure your workspace is stable and
free of any air currents or vibration. In some cases,
nearby radio wave emitting devices can cause
fluctuations.