background image

I.  Part List

1.  Scale

2.  Cover/Expansion Tray (top and bottom)

3.  AAA Batteries (x2)

II.  Key Description

1.  Power Key

 - Turns the scale on or off.

2.  Mode Key

 - Changes the displayed unit of measure. 

Enters calibration if pressed during startup (see VII)

3.  Tare Key

 - Re-zeros the scale. Used to remove the 

weight of a tray or container.

4.  Backlight Key

 - Turns backlight on or off.

III. Battery Installation

Proper Battery Installation

1.

4.

3.

2.

g

Summary of Contents for BLADE V2 BL2-50

Page 1: ...American Weigh Scales BL2 50 50g x 0 01g BL2 100 100g x 0 01g BL2 400 400g x 0 1g User Manual BLADE V2 Copyright 2011 American Weigh Scales Inc All rights reserved Rev 2 0 ...

Page 2: ...r 30 60 seconds before performing calibration if available Store your scale in a clean dry location Dust dirt and moisture can accumulate on the weighing sensors and electronics causing inaccuracy or malfunction Avoid using your scale in close proximity to strong radio wave emitting devices such as cordless phones Always weigh on a flat and level surface free from vibrations and drafts The corner ...

Page 3: ...urns the scale on or off 2 Mode Key Changes the displayed unit of measure Enters calibration if pressed during startup see VII 3 Tare Key Re zeros the scale Used to remove the weight of a tray or container 4 Backlight Key Turns backlight on or off III Battery Installation Proper Battery Installation 1 4 3 2 g ...

Page 4: ... Place the scale on a flat level surface and press the key to turn the scale on 2 Once the start up process completes and the scale goes to zero you are ready to weigh 3 Place the quantity to be weighed onto the weighing platform 4 Wait a couple seconds for the display to stabilize before taking a weight reading 5 To turn the scale OFF press the key again V Changing the Unit of Measure 1 Press the...

Page 5: ...ight of the contents 4 To return to normal weight display gross weight remove any items from the weighing platform and press the T key again to re zero the scale VII Calibration Initial calibration is performed at the factory However the calibration can drift slightly due to stresses from shipping Also due to changes in local acceleration of gravity around the world adjustment is recommended when ...

Page 6: ...ibration weight see above 3 Place the required calibration weight on the center of the weighing platform 4 Once the weight is confirmed the display will show PASS then return to normal weighing mode NOTE Adjustment using the incorrect mass will result in inaccurate readings IX Troubleshooting Problem Scale will not turn on Solution Check the batteries and battery contacts Problem Displayed weight ...

Page 7: ...ts the load cell may be permanently damaged due to overloading 2 O Ld Overload The maximum capacity has been exceeded Remove the extra weight to avoid permanently damaging the load cell 3 Lo Batteries are low Replace the batteries XI Specifications BL2 50 BL2 100 BL2 400 Max Capacity 50g 100g 400g Readability 0 01g 0 01g 0 1g Min Weight 0 03g 0 03g 0 2g Power 2 x 1 5V AAA Alkaline Dimensions 2 9 x...

Page 8: ... Readability Charts BL2 50 BL2 100 50g x 0 01g 1 76oz x 0 001oz 1 61ozt x 0 001ozt 32 15dwt x 0 01dwt 250ct x 0 05ct 771 6gn x 0 2gn 100g x 0 01g 3 527oz x 0 001oz 3 215ozt x 0 001ozt 64 30dwt x 0 01dwt 500ct x 0 05ct 1543 2gn x 0 2gn BL2 400 400g x 0 1g 14 11oz x 0 01oz 12 86ozt x 0 01ozt 257 2dwt x 0 1dwt 2000ct x 0 5ct 6173gn x 2gn ...