background image

the contents.

4.  To return to normal weight display (gross weight), 

remove any items from the weighing platform and 

press the

 TARE

 key again to re-zero the scale.

VII.  Calibration

Calibration and adjustment are performed at the factory. 

However, the weighing range can shift slightly during 

shipping. Also, due to changes in local acceleration of gravity 

around the world, adjustment is recommended when moving 

the scale from one location to another. To check calibration:

1.  Place the required calibration weight on the scale 

and note the displayed value.

2.  If the displayed value is not within acceptable 

tolerance (see chart below), adjustment is required.

Model

Acceptable 

Tolerance (±)

Calibration 

Weight

Recommended 

Accuracy Class

BL-100

0.02g

100g

OIML M2

BL-650

0.2g

500g

OIML M2

BL-1KG

0.2g

500g

OIML M2

VIII.  Adjustment

1.  Press and hold the 

MODE

 key until the display 

shows “CAL”.

2.  Press the 

MODE

 key again. The display will flash 

“CAL” followed by the calibration weight value (see 

chart).

3.  Place the calibration weight on the center of the 

weighing platform. Make sure you are using the 

correct weight for your specific model.

Summary of Contents for Blade-100

Page 1: ...BLADE 100 100g x 0 01g BLADE 650 650g x 0 1g BLADE 1KG 1000g x0 1g User Manual BLADE Scale Copyright 2010 American Weigh Scales Inc All rights reserved Rev 3 1 US Patent D525 891...

Page 2: ...ation and adjustment if available Store your scale in a clean dry location Dust dirt and moisture can accumulate on the weighing sensors and electronics causing inaccuracy or malfunction Avoid using y...

Page 3: ...re using the scale for the first time check the battery compartment and remove any plastic insulation strips or packaging that may be in place to prevent battery drainage during shipping storage 2 If...

Page 4: ...will default to the last unit used BLADE 100 Units g gram oz ounce ct carat gn grain BLADE 650 BLADE 1KG Units g gram oz ounce ozt troy ounce dwt pennyweight VI Using the Tare Feature You can use thi...

Page 5: ...ration 1 Place the required calibration weight on the scale and note the displayed value 2 If the displayed value is not within acceptable tolerance see chart below adjustment is required Model Accept...

Page 6: ...e your workspace is stable and free of any air currents or vibration Problem Displayed weight is inaccurate Solution Perform a calibration adjustment X Error Codes 1 OUTZ Zero range has shifted Perfor...

Page 7: ...y 650g 1000g Readability 0 1g 0 1g Min Weight 0 2g 0 2g Power 2 x AAA Dimensions 0 8 x 3 8 x 3 8 Platform Dimensions 2 7 x 2 7 XII Capacity Readability Charts BL 100 100g x 0 01g 3 528oz x 0 001oz 500...